Après circulation extra-corporelle, la mesure continue de la Svo2 et la co-oxymétrie du sang mêlé ne sont pas interchangeables
Authors:
- 1 June 2004
- journal article
- abstracts
- Published by Springer Nature in Canadian Journal of Anesthesia/Journal canadien d'anesthésie
- Vol. 51 (6) , 610-615
Abstract
Déterminer la précision de la mesure continue (in vivo) de la saturation veineuse en oxygène du sang mêlé (Svo2) par cathéter à fibres optiques chez les patients en post-circulation extracorporelle (CEC). Nous avons étudié prospectivement 14 patients (âge, 64 ± 8) post-CEC équipés d’un cathéter artériel pulmonaire. Le taux d’hématocrite moyen était de 30 ± 4 %. Le cathéter a été calibréin vitro etin vivo suivant les recommandations du constructeur. Cinquante-six mesures simultanées de Svo2 continue (CSvo2) et de Svo2 mesurée (MSvo2) par hémoxymètre ont été effectuées et les valeurs pairées ont été analysées par régression linéaire. Pour que les deux mesures soient interchangeables, nous avons déterminé,a priori, une limite maximale de 3% (environ 5% de la mesure) comme différence acceptable entre les deux types de mesures. Toutes les mesures ont été faites dans les quatre heures suivant la mise en place du cathéter. CSvo2 est faiblement corrélée à MSvo2 avec un coefficient de corrélation r2 = 0,49 (P < 0,001). L’analyse de Bland-Altman objective un biais moyen de 0,8 ± 3% avec 36% des valeurs mesurées en dehors de la zone cliniquement acceptable. Pour des valeurs CSvo2 ± 65%, les mesures ne sont pas corrélées (P = NS), le biais est de -2 ± 7% et 56% des valeurs mesurées en dehors de la zone cliniquement acceptable. Nos résultats objectivent que, dans les suites immédiates d’une CEC, la mesure continue de la Svo2 et la co-oxymétrie du sang mêlé ne sont pas interchangeables. D’autres études sont nécessaires afin de confirmer ces résultats et comprendre les phénomènes physiopathologiques a l’origine de l’absence de concordance entre les deux méthodes. To determine the accuracy of continuous (in vivo) measurement of mixed venous oxygen saturation (SvO2), using a fibreoptic catheter, in patients having had cardiopulmonary bypass (CPB). Using a pulmonary arterial catheter, we prospectively studied 14 patients (age 64 ± 8) having had cardiopulmonary bypass. Mean hematocrit was 30 ± 4%. The catheter was calibrated in vitro and in vivo, according to the manufacturer’s instructions. Fifty-six simultaneous measurements of continuous SvO2 (CSvO2) and measured SvO2 (MSvO2) were taken with a co-oxymeter and the paired values were analyzed by the linear regression method. To make the two sets of measurements interchangeable, we established, a priori, a maximum limit of 3% (approximately 5% of the measurement), as being an acceptable difference between the two types of measurements. All the measurements were obtained within four hours of the placement of the catheter. CSvO2 was weakly correlated with MSvO2, with a correlation coefficient of r2 = 0.49 (P < 0.001). The Bland-Altman analysis demonstrates an objective mean bias of 0.8 ± 3%, with 36% of the values measured falling outside clinically acceptable limits. For values of CSvO2 ± 65%, the measurements were not correlated (P = NS). The bias was −2 ± 7%, and 56% of the values measured fell outside clinically acceptable limits. Our results demonstrate that, in the immediate aftermath of cardiopulmonary bypass, the continuous measurement of SvO2 and the co-oxymetry determination of mixed blood is not interchangeable. Other studies will be necessary to confirm these results and enable us to understand the pathophysiological phenomena accounting for the lack of agreement between the two methods.Keywords
This publication has 13 references indexed in Scilit:
- Perioperative evaluation of a new mixed venous oxygen saturation catheter in cardiac surgical patientsJournal of Cardiothoracic and Vascular Anesthesia, 1992
- A bedside index assessing the reliability of pulmonary artery occlusion pressure measurements during mechanical ventilation with positive end-expiratory pressureJournal of Critical Care, 1992
- Continuous monitoring of mixed venous oxygen saturation during aortic operationsCritical Care Medicine, 1992
- Accuracy of a mixed venous saturation catheter during acutely induced changes in hematocrit in humansCritical Care Medicine, 1991
- Quelle est la place du monitorage de la saturation du sang veineux mêlé en chirurgie cardiaque?Annales Françaises d'Anesthésie et de Réanimation, 1989
- Continuous Venous Oximetry in Surgical PatientsAnnals of Surgery, 1986
- Value of continuous monitoring of mixed venous blood oxygen saturation in the management of critically ill patientsCritical Care Medicine, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Continuous Monitoring of Mixed Venous Oxygen SaturationChest, 1984
- Controlled Sedation with Alphaxalone-AlphadoloneBMJ, 1974