Eddy‐current compensated diffusion weighting with a single refocusing RF pulse
Open Access
- 8 January 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 61 (3) , 748-754
Abstract
A modification of the Stejskal‐Tanner diffusion‐weighting preparation with a single refocusing RF pulse is presented which involves three gradient lobes that can be adjusted to null eddy currents with any given decay rate to reduce geometric distortions in diffusion‐weighted echo‐planar imaging (EPI). It has a very similar compensation performance as the commonly used double‐spin‐echo preparation but (i) is less sensitive to flip angle imperfections, e.g. along the slice profile, and B1 inhomogeneities and (ii) can yield shorter echo times for moderate b values, notably for longer echo trains as required for higher spatial resolution. It therefore can provide an increased signal‐to‐noise ratio as is simulated numerically and demonstrated experimentally in water phantoms and the human brain for standard EPI (2.0 × 2.0 mm2) and high‐resolution EPI of inner field‐of‐views using 2D‐selective RF excitations (0.5 × 1.0 mm2). Thus, the presented preparation may help to overcome current limitations of diffusion‐weighted EPI, in particular at high static magnetic fields. Magn Reson Med, 2009.Keywords
This publication has 19 references indexed in Scilit:
- DWI of the spinal cord with reduced FOV single‐shot EPIMagnetic Resonance in Medicine, 2008
- Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneityMagnetic Resonance in Medicine, 2006
- Echoplanar diffusion tensor imaging of the lower leg musculature using eddy current nulled stimulated echo preparationMagnetic Resonance in Medicine, 2006
- Eddy current correction in diffusion‐weighted imaging using pairs of images acquired with opposite diffusion gradient polarityMagnetic Resonance in Medicine, 2003
- Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imagingMagnetic Resonance in Medicine, 2002
- Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systemsMagnetic Resonance in Medicine, 1991
- Optimization of eddy-current compensationJournal of Magnetic Resonance (1969), 1990
- A k-space analysis of small-tip-angle excitationJournal of Magnetic Resonance (1969), 1989
- Multishield active magnetic screening of coil structures in NMRJournal of Magnetic Resonance (1969), 1987
- Active magnetic screening of gradient coils in NMR imagingJournal of Magnetic Resonance (1969), 1986