Clinical CT-based calculations of dose and positron emitter distributions in proton therapy using the FLUKA Monte Carlo code
Authors:
- 17 May 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (12) , 3369-3387
Abstract
Clinical investigations on post-irradiation PET/CT (positron emission tomography/computed tomography) imaging for in vivo verification of treatment delivery and, in particular, beam range in proton therapy are underway at Massachusetts General Hospital (MGH). Within this project, we have developed a Monte Carlo framework for CT-based calculation of dose and irradiation-induced positron emitter distributions. Initial proton beam information is provided by a separate Geant4 Monte Carlo simulation modelling the treatment head. Particle transport in the patient is performed in the CT voxel geometry using the FLUKA Monte Carlo code. The implementation uses a discrete number of different tissue types with composition and mean density deduced from the CT scan. Scaling factors are introduced to account for the continuous Hounsfield unit dependence of the mass density and of the relative stopping power ratio to water used by the treatment planning system (XiO (Computerized Medical Systems Inc.)). Resulting Monte Carlo dose distributions are generally found in good correspondence with calculations of the treatment planning program, except a few cases (e.g. in the presence of air/tissue interfaces). Whereas dose is computed using standard FLUKA utilities, positron emitter distributions are calculated by internally combining proton fluence with experimental and evaluated cross-sections yielding 11C, 15O, 14O, 13N, 38K and 30P. Simulated positron emitter distributions yield PET images in good agreement with measurements. In this paper, we describe in detail the specific implementation of the FLUKA calculation framework, which may be easily adapted to handle arbitrary phase spaces of proton beams delivered by other facilities or include more reaction channels based on additional cross-section data. Further, we demonstrate the effects of different acquisition time regimes (e.g., PET imaging during or after irradiation) on the intensity and spatial distribution of the irradiation-induced beta+-activity signal for the cases of head and neck and para-spinal tumour sites.Keywords
This publication has 26 references indexed in Scilit:
- Patient Study of In Vivo Verification of Beam Delivery and Range, Using Positron Emission Tomography and Computed Tomography Imaging After Proton TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- PET/CT imaging for treatment verification after proton therapy: A study with plastic phantoms and metallic implantsMedical Physics, 2007
- Conversion of CT numbers into tissue parameters for Monte Carlo dose calculations: a multi-centre studyPhysics in Medicine & Biology, 2007
- Adaptation of GEANT4 to Monte Carlo dose calculations based on CT dataMedical Physics, 2004
- Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facilityMedical Physics, 2004
- A pencil beam algorithm for proton dose calculationsPhysics in Medicine & Biology, 1996
- Proton dose monitoring with PET: quantitative studies in LucitePhysics in Medicine & Biology, 1996
- Proton therapy in combination with PET as monitor: a feasibility studyIEEE Transactions on Nuclear Science, 1993
- Production ofradioactive nuclei from,n) or(p,α) reactionsPhysical Review C, 1989
- total nuclear cross sectionPhysical Review C, 1978