Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
- 1 January 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Microwave Theory and Techniques
- Vol. 67 (3) , 1265-1273
Abstract
Patients with implanted medical devices such as deep brain stimulation or spinal cord stimulation are often unable to receive magnetic resonance imaging (MRI). This is because, once the device is within the radio frequency (RF) field of the MRI scanner, electrically conductive leads act as antenna, amplifying the RF energy deposition in the tissue and causing possible excessive tissue heating. Here, we propose a novel concept in lead design in which 40-cm lead wires are coated with a ~1.2-mm layer of high dielectric constant material (155 <; ε r <; 250) embedded in a weakly conductive insulation (σ = 20 S/m). The technique called high-dielectric capacitive bleeding of current (CBLOC) works by forming a distributed capacitance along the lengths of the lead, efficiently dissipating RF energy before it reaches the exposed tip. Measurements during RF exposure at 64 and 123 MHz demonstrated that CBLOC leads generated 20-fold less heating at 1.5 T and 40-fold less heating at 3 T compared to control leads. Numerical simulations of RF exposure at 297 MHz (7 T) predicted a 15-fold reduction in specific absorption rate of RF energy around the tip of CBLOC leads compared to control leads.Keywords
All Related Versions
Funding Information
- National Institutes of Health (R03EB024705)
This publication has 51 references indexed in Scilit:
- Analysis of fractal electrodes for efficient neural stimulationFrontiers in Neuroengineering, 2013
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- EFFECT OF REALISTIC MODELING OF DEEP BRAIN STIMULATION ON THE PREDICTION OF VOLUME OF ACTIVATED TISSUEProgress In Electromagnetics Research, 2012
- Neuroimaging and Deep Brain StimulationAmerican Journal of Neuroradiology, 2009
- Measuring RF‐induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leadsMagnetic Resonance in Medicine, 2009
- Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurationsBioMedical Engineering OnLine, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- Investigating the effects of external fields polarization on the coupling of pure magnetic waves in the human body in very low frequenciesBioMagnetic Research and Technology, 2007
- Material characterization of a high‐dielectric‐constant polymer–ceramic composite for embedded capacitor for RF applicationsJournal of Applied Polymer Science, 2004
- RF heating of implanted spinal fusion stimulator during magnetic resonance imagingIEEE Transactions on Biomedical Engineering, 1997