Calibration and validation of a quality assurance system for90Sr/90Y radiation source trains
Authors:
- P RosenthalPRP Rosenthal
- W WeberWWW Weber
- A F RsterARA F Rster
- O OrthOOO Orth
- B K HlerBHB K Hler
- F SeilerFSF Seiler
- 17 February 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (5) , 573-585
Abstract
A quality assurance system (OPTIDOS, PTW-Freiburg) developed for dose rate verification of 90Sr/90Y radiation source trains (RSTs) was calibrated and validated. These source trains are used in the 5-F-BetaCath system (Novoste Corp.) for the treatment of endovascular diseases. The calibration factor of the OPTIDOS system was obtained empirically and is valid for 90Sr/90Y dose rate measurements at the specification point which is located at 2 mm distance from the source axis. A total of 187 OPTIDOS dose rate verifications of the 5-F-BetaCath system were performed in different hospitals. The histogram of the deviation between the manufacturer's dose rate specification and the dose rate measured using the OPTIDOS dosimetry system is Gaussian shaped with +/- 3% relative width and a mean shift of about +2% with respect to the corresponding dose rate specification. Additionally, 128 OPTIDOS dose rate verifications of the new jacketed RST (3.5-F-BetaCath, Novoste Corp.) were performed using the same calibration factor as derived for the 5-F-BetaCath system. Distribution of the deviation between the certified and the measured dose rate is nearly identical in comparison to the histogram of the 5-F-BetaCath system. The mean value of the deviations is shifted by -1.5% with respect to the certified dose rate. In order to compare the results of the calibrated OPTIDOS dosimetry system with a standard measuring method, separate dose rate measurements were performed using electron accelerator calibrated radiochromic films in which calibration is traceable to PTB (Physikalisch Technische Bundesanstalt, Germany). Deviation between both the methods is less than 3.1%. These results confirm that the calibrated OPTIDOS dosimetry system can be considered suitable for quality assurance of both types of RST used in the BetaCath systems.Keywords
This publication has 6 references indexed in Scilit:
- Clinical and angiographic outcomes after use of 90Strontium/90Yttrium beta radiation for the treatment of in-stent restenosis: results from the Stents and Radiation Therapy 40 (START 40) registryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapyMedical Physics, 2000
- Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60Medical Physics, 1999
- Direct reading measurement of absorbed dose with plastic scintillators—The general concept and applications to ophthalmic plaque dosimetryMedical Physics, 1996
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991
- Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991