Ultrasonography-based 2D motion-compensated HIFU sonication integrated with reference-free MR temperature monitoring: a feasibility studyex vivo
Authors:
- Lorena PetruscaLPLorena Petrusca
- Magalie ViallonMagalie Viallon
- Thomas GogetTGThomas Goget
- Christoph D BeckerCBChristoph D Becker
- Rares SalomirRSRares Salomir
- 20 April 2012
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (10) , N159-N171
Abstract
Magnetic resonance imaging (MRI) and ultrasonography have been used simultaneously in this ex vivo study for the image-guidance of high intensity focused ultrasound (HIFU) treatment in moving tissue. A ventilator-driven balloon produced periodic and non-rigid (i.e. breathing-like) motion patterns in phantoms. MR-compatible ultrasound (US) imaging enabled near real-time 2D motion tracking based on optical flow detection, while near-harmonic reference-free proton resonance frequency shift (PRFS) MR thermometry (MRT) was used to monitor the thermal buildup on line. Reference-free MRT was applied to gradient-echo echo-planar imaging phase maps acquired at the frame rate of 250 to 300 ms/slice with voxel size 1.25×1.25×5 mm(3). The MR-US simultaneous imaging was completely free of mutual interferences while minor RF interferences from the HIFU device were detected in the far field of the US images. The effective duty-cycle of the HIFU sonication was close to 100 % and no off-interval was required to temporally decouple it from the ultrasonography. The motion compensation of the HIFU sonication was achieved with an 8 Hz frame rate and sub-millimeter spatial accuracy, both for single-focus mode and for an iterated multi-foci line scan. Near harmonic reference-less PRFS MRT delivered motion-robust thermal maps perpendicular or parallel to the HIFU beam (0.7 °C precision, 0.5 °C absolute accuracy). Out-of-plane motion compensation was not addressed in this study.Keywords
This publication has 36 references indexed in Scilit:
- An MR-compliant phased-array HIFU transducer with augmented steering range, dedicated to abdominal thermotherapyPhysics in Medicine & Biology, 2011
- Rapid motion correction in MR‐guided high‐intensity focused ultrasound heating using real‐time ultrasound echo informationNMR in Biomedicine, 2010
- Hybrid ultrasound MRI for improved cardiac imaging and real‐time respiration controlMagnetic Resonance in Medicine, 2009
- Thermal ablation of uterine fibroids using MR-guided focused ultrasound-a truly non-invasive treatment modalityEuropean Radiology, 2007
- Real‐time adaptive methods for treatment of mobile organs by MRI‐controlled high‐intensity focused ultrasoundMagnetic Resonance in Medicine, 2007
- The evolving non-surgical ablation of breast cancer: Mr Guided focused ultrasound (MRgFUS)Breast Cancer, 2007
- MR-guided focused ultrasound surgery (MRgFUS) for the palliation of pain in patients with bone metastases—preliminary clinical experienceAnnals of Oncology, 2006
- MR-Guided Focused Ultrasound SurgeryJournal of Computer Assisted Tomography, 1992
- Rapid NMR Cardiography with a Half-Echo M-Mode MethodJournal of Computer Assisted Tomography, 1991
- Production of Focal Destructive Lesions in the Central Nervous System With UltrasoundJournal of Neurosurgery, 1954