Postal intercomparison of absorbed dose for high energy x rays with thermoluminescence dosimeters

Authors:
  • BBB. E. Bjärngard
    B. E. Bjärngard
  • KKK. R. Kase
    K. R. Kase
  • BRB.-I. Rudén
    B.-I. Rudén
  • PBP. J. Biggs
    P. J. Biggs
  • ABA. L. Boyer
    A. L. Boyer
  • KJK. A. Johansson
    K. A. Johansson

Abstract
The accuracy and precision of the IAEA/WHO [International Atomic Energy Agency/ World Health Organization] LiF TLD [thermoluminescent dosimeter] system used in intercomparison by mail of absorbed doses from 60Co .gamma.-radiation and 4-25 MV X-rays was studied. The system employs 160 mg LiF powder in polystyrene capsules, which are placed at 5 or 7 cm depth in water and irradiated to doses close to 200 rad (2.00 Gy [gray]). The dosimeters are mailed to the IAEA Dosimetry Laboratory and read under conditions to minimize variations in instrument sensitivity. The precision of the readout technique, using 3 capsules/irradiation and the readout of 5 aliquots/capsule, is characterized by 0.2% SD of the resulting mean. Since random errors during the irradiation are added, the detectable systematic discrepancy in dose delivery at the 95% confidence level is .+-. 2% for 60Co .gamma. and .+-. 3% for high-energy X-rays.
Funding Information
  • IAEA Research