Failure mode and effects analysis based risk profile assessment for stereotactic radiosurgery programs at three cancer centers in Brazil
- 22 December 2015
- journal article
- Published by Wiley in Medical Physics
- Vol. 43 (1) , 171-178
Abstract
Purpose: The goal of this study was to evaluate the safety and quality management program for stereotactic radiosurgery(SRS) treatment processes at three radiotherapy centers in Brazil by using three industrial engineering tools (1) process mapping, (2) failure modes and effects analysis (FMEA), and (3) fault tree analysis. Methods: The recommendations of Task Group 100 of American Association of Physicists in Medicine were followed to apply the three tools described above to create a process tree for SRS procedure for each radiotherapy center and then FMEA was performed. Failure modes were identified for all process steps and values of risk priority number (RPN) were calculated from O, S, and D (RPN = O × S × D) values assigned by a professional team responsible for patient care. Results: The subprocess treatment planning was presented with the highest number of failure modes for all centers. The total number of failure modes were 135, 104, and 131 for centers I, II, and III, respectively. The highest RPN value for each center is as follows: center I (204), center II (372), and center III (370). Failure modes with RPN ≥ 100: center I (22), center II (115), and center III (110). Failure modes characterized by S ≥ 7, represented 68% of the failure modes for center III, 62% for center II, and 45% for center I. Failure modes with RPNs values ≥100 and S ≥ 7, D ≥ 5, and O ≥ 5 were considered as high priority in this study. Conclusions: The results of the present study show that the safety risk profiles for the same stereotactic radiotherapy process are different at three radiotherapy centers in Brazil. Although this is the same treatment process, this present study showed that the risk priority is different and it will lead to implementation of different safety interventions among the centers. Therefore, the current practice of applying universal device-centric QA is not adequate to address all possible failures in clinical processes at different radiotherapy centers. Integrated approaches to device-centric and process specific quality management program specific to each radiotherapy center are the key to a safe quality management program.Keywords
This publication has 11 references indexed in Scilit:
- Application of failure mode and effects analysis to intracranial stereotactic radiation surgery by linear accelerator In Regard to Masini et alPractical Radiation Oncology, 2015
- Application of failure mode and effects analysis to intracranial stereotactic radiation surgery by linear acceleratorPractical Radiation Oncology, 2014
- Process‐based quality management for clinical implementation of adaptive radiotherapyMedical Physics, 2014
- A streamlined failure mode and effects analysisMedical Physics, 2014
- Quality Assurance With Plan Veto: Reincarnation of a Record and Verify System and Its Potential ValueInternational Journal of Radiation Oncology, Biology, Physics, 2014
- The Use of Failure Mode and Effect Analysis in a Radiation Oncology Setting: The Cancer Treatment Centers of America ExperienceJournal for Healthcare Quality, 2014
- Failure Mode and Effect Analysis for Delivery of Lung Stereotactic Body Radiation TherapyInternational Journal of Radiation Oncology, Biology, Physics, 2012
- Application of Failure Mode and Effects Analysis to Intraoperative Radiation Therapy Using Mobile Electron Linear AcceleratorsInternational Journal of Radiation Oncology, Biology, Physics, 2012
- Applying failure mode effects and criticality analysis in radiotherapy: Lessons learned and perspectives of enhancementRadiotherapy and Oncology, 2010
- Evaluation of Safety in a Radiation Oncology Setting Using Failure Mode and Effects AnalysisInternational Journal of Radiation Oncology, Biology, Physics, 2009