Abstract
Highly sensitive and selective nonenzymatic detection of glucose has been achieved using a novel disposable electrochemical sensor based on three-dimensional (3D) porous nickel nanostructures. The enzyme-free sensor was fabricated through in situ growing porous nickel networks on a homemade screen-printed carbon electrode substrate via electrochemically reducing the Ni2+ precursor, along with continuously liberating hydrogen bubbles. The resulting nickel-modified electrode was characterized by scanning electron microscopy (SEM), transmission electron microscopy (TEM), energy-dispersive X-ray spectrometry (EDX), powder X-ray diffractometry (XRD), and electrochemical techniques. Cyclic voltammetric, alternating-current impedance, and amperometric methods were used to investigate the catalytic properties of the assembled sensor for glucose electro-oxidation in alkaline media. Under optimized conditions, the enzymeless sensor exhibited excellent performance for glucose analysis selectively, offering a much wider linear range (from 0.5 μM to 4 mM), an extremely low detection limit (0.07 μM, signal-to-noise ratio (S/N) of 3), and an ultrahigh sensitivity of 2.9 mA/(cm2 mM). Importantly, favorable reproducibility and long-term performance stability were obtained thanks to the robust frameworks. Application of the proposed sensor in monitoring blood glucose was also demonstrated.