Auditory Force Feedback Substitution Improves Surgical Precision during Simulated Ophthalmic Surgery
- 1 February 2013
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 54 (2) , 1316-1324
Abstract
To determine the extent that auditory force feedback (AFF) substitution improves performance during a simulated ophthalmic peeling procedure. A 25-gauge force-sensing microforceps was linked to two AFF modes. The "alarm" AFF mode sounded when the force reached 9 mN. The "warning" AFF mode made beeps with a frequency proportional to the generated force. Participants with different surgical experience levels were asked to peel a series of bandage strips off a platform as quickly as possible without exceeding 9 mN of force. In study arm A, participants peeled with alarm and warning AFF modes, the order randomized within the experience level. In study arm B, participants first peeled without AFF, then alarm or warning AFF (order randomized within the experience level), and finally without AFF. Of the 28 "surgeon" participants, AFF improved membrane peeling performance, reducing average force generated (P < 0.01), SD of forces (P < 0.05), and force × time above 9 mN (P < 0.01). Short training periods with AFF improved subsequent peeling performance when AFF was turned off, with reductions in average force, SD of force, maximum force, time spent above 9 mN, and force × time above 9 mN (all P < 0.001). Except for maximum force, peeling with AFF reduced all force parameters (P < 0.05) more than peeling without AFF after completing a training session. AFF enables the surgeon to reduce the forces generated with improved precision during phantom membrane peeling, regardless of surgical experience. New force-sensing surgical tools combined with AFF offer the potential to enhance surgical training and improve surgical performance.Keywords
This publication has 13 references indexed in Scilit:
- A FORCE-SENSING MICROSURGICAL INSTRUMENT THAT DETECTS FORCES BELOW HUMAN TACTILE SENSATIONRetina, 2013
- Force sensing micro-forceps for robot assisted retinal surgery2012
- Miniature fiber-optic force sensor based on low-coherence Fabry-Pérot interferometry for vitreoretinal microsurgeryBiomedical Optics Express, 2012
- Visual force feedback in laparoscopic trainingSurgical Endoscopy, 2011
- The Role of Haptic Feedback in Laparoscopic Simulation TrainingJournal of Surgical Research, 2009
- A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgeryInternational Journal of Computer Assisted Radiology and Surgery, 2009
- The value of haptic feedback in conventional and robot-assisted minimal invasive surgery and virtual reality training: a current reviewSurgical Endoscopy, 2009
- Effects of visual force feedback on robot-assisted surgical task performanceThe Journal of Thoracic and Cardiovascular Surgery, 2008
- Applied force during vitreoretinal microsurgery with handheld instruments2007
- Application of Haptic Feedback to Robotic SurgeryJournal of Laparoendoscopic & Advanced Surgical Techniques, 2004