A fully sealed plastic chip for multiplex PCR and its application in bacteria identification
Authors:
- 1 May 2015
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 15 (13) , 2826-2834
Abstract
Multiplex PCR is an effective tool for simultaneous multiple target detection but is limited by the intrinsic interference and competition among primer pairs when it is performed in one reaction tube. Dividing a multiplex PCR into many single PCRs is a simple strategy to overcome this issue. Here, we constructed a plastic, easy-to-use, fully sealed multiplex PCR chip based on reversible centrifugation for the simultaneous detection of 63 target DNA sequences. The structure of the chip is quite simple, which contains sine-shaped infusing channels and a number of reaction chambers connecting to one side of these channels. Primer pairs for multiplex PCR were sequentially preloaded in the different reaction chambers, and the chip was enclosed with PCR-compatible adhesive tape. For usage, the PCR master mix containing a DNA template is pipetted into the infusing channels and centrifuged into the reaction chambers, leaving the infusing channels filled with air to avoid cross-contamination of the different chambers. Then, the chip is sealed and placed on a flat thermal cycler for PCR. Finally, amplification products can be detected in situ using a fluorescence scanner or recovered by reverse centrifugation for further analyses. Therefore, our chip possesses two functions: 1) it can be used for multi-target detection based on end-point in situ fluorescence detection; and 2) it can work as a sample preparation unit for analyses that need multiplex PCR such as hybridization and target sequencing. The performance of this chip was carefully examined and further illustrated in the identification of 8 pathogenic bacterial genomic DNA samples and 13 drug-resistance genes. Due to simplicity of its structure and operation, accuracy and generality, high-throughput capacity, and versatile functions (i.e., for in situ detection and sample preparation), our multiplex PCR chip has great potential in clinical diagnostics and nucleic acid-based point-of-care testing.Keywords
This publication has 36 references indexed in Scilit:
- 30.Journal of Clinical Neuroscience, 2014
- Development of a new 26plex Y-STRs typing system for forensic applicationForensic Science International: Genetics, 2014
- Microfluidic approaches for cancer cell detection, characterization, and separationLab on a Chip, 2012
- Food analysis and food authentication by peptide nucleic acid (PNA)-based technologiesChemical Society Reviews, 2010
- Rapid detection and identification of 12 respiratory viruses using a dual priming oligonucleotide system-based multiplex PCR assayJournal of Virological Methods, 2009
- Large scale multiplex PCR improves pathogen detection by DNA microarraysBMC Microbiology, 2009
- Multiple-locus variable-number tandem-repeats analysis of Salmonella enterica subsp. enterica serovar Typhimurium using PCR multiplexing and multicolor capillary electrophoresisJournal of Microbiological Methods, 2004
- Solid phase DNA amplification: characterisation of primer attachment and amplification mechanismsNucleic Acids Research, 2000
- Multiplex PCR: Optimization and Application in Diagnostic VirologyClinical Microbiology Reviews, 2000
- Deletion screening of the Duchenne muscular dystrophy locus via multiplex DNA amplificationNucleic Acids Research, 1988