Rapid liquid chromatography–tandem mass spectrometry method for quantification of ziprasidone in human plasma

Authors:

Abstract
A liquid chromatography–tandem mass spectrometry (LC‐MS/MS) method for the determination of ziprasidone (ZIP) in human plasma was developed. ZIP and N‐methyl ziprasidone as internal standard (IS) were extracted from alkalinized plasma using tert‐ butyl methyl ether. Separation was performed isocratically on a C8 column with 90% acetonitrile containing 2 mmol/L ammonium acetate as a mobile phase with a total run time of 2.5 min. MS/MS transitions of m/z 413 → 194 and m/z 427 → 177 of the analyte and internal standard were used for quantification. Confirmatory ions of m/z 413 → 177 and m/z 427 → 180 were collected as well. The calibration curve based on peak‐area ratio was linear up to at least 200 ng/mL with a detection limit of 0.1 ng/mL. The method showed satisfactory reproducibility with a coefficient of variation of less than 5%. The method was successfully applied to the analysis of ZIP in spiked human plasma. Copyright © 2005 John Wiley & Sons, Ltd.

This publication has 10 references indexed in Scilit: