Nitric Oxide-Releasing Hemodialysis Catheter Lock Solutions
- 7 June 2023
- journal article
- research article
- Published by American Chemical Society (ACS) in ACS Applied Materials & Interfaces
- Vol. 15 (24) , 28907-28921
Abstract
In an attempt to address the significant morbidity, mortality, and economic cost associated with tunneled dialysis catheter (TDC) dysfunction, we report the development of nitric oxide-releasing dialysis catheter lock solutions. Catheter lock solutions with a range of NO payloads and release kinetics were prepared using low-molecular-weight N-diazeniumdiolate nitric oxide donors. Nitric oxide released through the catheter surface as a dissolved gas was maintained at therapeutically relevant levels for at least 72 h, supporting clinical translatability (interdialytic period). Slow, sustained NO release from the catheter surface prevented bacterial adhesion in vitro by 88.9 and 99.7% for Pseudomonas aeruginosa and Staphylococcus epidermidis, respectively, outperforming a burst NO-release profile. Furthermore, bacteria adhered to the catheter surface in vitro prior to lock solution use was reduced by 98.7 and 99.2% for P. aeruginosa and S. epidermidis, respectively, when using a slow releasing NO donor, demonstrating both preventative and treatment potential. The adhesion of proteins to the catheter surface, a process often preceding biofilm formation and thrombosis, was also lessened by 60-65% by sustained NO release. In vitro cytotoxicity of catheter extract solutions to mammalian cells was minimal, supporting the non-toxic nature of the NO-releasing lock solutions. The use of the NO-releasing lock solution in an in vivo TDC porcine model demonstrated decreased infection and thrombosis, enhanced catheter functionality, and improved outcome (i.e., likelihood of survival) as a result of catheter use.Keywords
Funding Information
- Division of Chemistry (CHE-0922858)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01DK108318)
- Eshelman Innovation
- KnowBio, LLC
- UNC Kidney Center
This publication has 66 references indexed in Scilit:
- Surface Modification of Silicone for Biomedical Applications Requiring Long-Term Antibacterial, Antifouling, and Hemocompatible PropertiesLangmuir, 2012
- Examination of bacterial resistance to exogenous nitric oxideNitric Oxide, 2012
- Nitric oxide release: Part II. Therapeutic applicationsChemical Society Reviews, 2012
- Nitric oxide release: Part I. Macromolecular scaffoldsChemical Society Reviews, 2012
- Nitric Oxide-Generating Silicone As a Blood-Contacting BiomaterialASAIO Journal, 2011
- Fifty Years of Diazeniumdiolate Research. From Laboratory Curiosity to Broad-Spectrum Biomedical AdvancesACS Chemical Biology, 2011
- Sodium Citrate Versus Heparin Catheter Locks for Cuffed Central Venous Catheters: A Single-Center Randomized Controlled TrialAmerican Journal of Kidney Diseases, 2009
- Reduced bacterial adhesion to fibrinogen-coated substrates via nitric oxide releaseBiomaterials, 2008
- Bactericidal Efficacy of Nitric Oxide-Releasing Silica NanoparticlesACS Nano, 2008
- The hemodialysis catheter conundrum: Hate living with them, but can’t live without themKidney International, 1999