A detailed process map for clinical workflow of a new biology‐guided radiotherapy (BgRT) machine
Authors:
Open Access
- 10 May 2022
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (6) , e13606
Abstract
Purpose: Biology‐guided radiotherapy (BgRT) is a new external beam radiation therapy modality combining PET‐CT with a linear accelerator that has the potential to track and treat one or more tumors in real‐time. The use of PET and radiopharmaceutical tracers introduces new processes that are different from the existing treatment processes. In this study, we have developed a process map for the clinical implementation of a prototype BgRT machine.Methods: A team of 13 members from various radiation therapy disciplines at our institution participated in developing a prospective process map for a prototype BgRT machine. The methodology provided by the AAPM TG 100 report was followed. In particular, the steps unique to the BgRT workflow, using hypofractionated stereotactic body radiation therapy with fluorodeoxyglucose radiolabeled with fluorine‐18 (FDG) to guide beam delivery, were analyzed.Results: The multi‐disciplinary team in the department of radiation oncology at our institution developed a prospective process map for the clinical BgRT workflow. By focusing on the appropriate level of detail, 15 major subprocesses, 133 steps, and 248 substeps were identified and the process map was agreed upon as being useful, implementable, and manageable. Seventy‐four steps from nine subprocesses, 55.6% of the whole process, were analyzed to be the BgRT unique steps. They originate mainly from: (1) acquiring multiple PET images at the BgRT machine with separate patient visits, (2) creating a unique biological treatment volume for BgRT plan (PTVBgRT), and (3) BgRT plan optimization and treatment delivery using PET images.Conclusion: Using BgRT to irradiate multiple metastases in the same session will impact clinical workflow, thus a graphical process map depicting the new clinical workflow with an appropriate level of detail is critical for efficient, safe, and high‐quality care. The prospective process map will guide the successful setup and use of the new BgRT system.Keywords
This publication has 13 references indexed in Scilit:
- Application of TG‐100 risk analysis methods to the acceptance testing and commissioning process of a Halcyon linear acceleratorMedical Physics, 2019
- Overall Survival with Durvalumab after Chemoradiotherapy in Stage III NSCLCNew England Journal of Medicine, 2018
- Stereotactic Ablative Radiation Therapy for the Comprehensive Treatment of Oligometastatic Tumors (SABR-COMET): Results of a Randomized TrialInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2018
- Consolidative Radiotherapy for Limited Metastatic Non–Small-Cell Lung CancerJAMA Oncology, 2017
- Local consolidative therapy versus maintenance therapy or observation for patients with oligometastatic non-small-cell lung cancer without progression after first-line systemic therapy: a multicentre, randomised, controlled, phase 2 studyThe Lancet Oncology, 2016
- The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality managementMedical Physics, 2016
- Toward a planning scheme for emission guided radiation therapy (EGRT): FDG based tumor tracking in a metastatic breast cancer patientMedical Physics, 2013
- Emission guided radiation therapy for lung and prostate cancers: A feasibility study on a digital patientMedical Physics, 2012
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010