The accuracy of clinical assessment techniques in diagnosing partial‐thickness burn injuries has remained as low as 50–76%. Depending on the burn depth and environmental factors in the wound, such as reactive oxygen species, inflammation, and autophagy, partial‐thickness burns can heal spontaneously or require surgical intervention. Herein, it is demonstrated that terahertz time‐domain spectral imaging (THz‐TDSI) is a promising tool for in vivo quantitative assessment and monitoring of partial‐thickness burn injuries in large animals. We used a novel handheld THz‐TDSI scanner to characterize burn injuries in a porcine scald model with histopathological controls. Statistical analysis (n = 40) indicates that the THz‐TDSI modality can accurately differentiate between partial‐thickness and full‐thickness burn injuries (1‐way ANOVA, p < 0.05). THz‐TDSI has the potential to improve burn care outcomes by helping surgeons in making objective decisions for early excision of the wound.Keywords
All Related Versions
Funding Information
- National Institute of General Medical Sciences (R01GM112693)
This publication has 57 references indexed in Scilit:
- Intense THz pulses cause H2AX phosphorylation and activate DNA damage response in human skin tissueBiomedical Optics Express, 2013
- In vivo terahertz imaging of rat skin burnsJournal of Biomedical Optics, 2012
- Terahertz reflectometry of burn wounds in a rat modelBiomedical Optics Express, 2011
- Dissecting the THz spectrum of liquid water from first principles via correlations in time and spaceProceedings of the National Academy of Sciences, 2010
- A porcine deep dermal partial thickness burn model with hypertrophic scarringBurns, 2006
- Meta-analysis of early excision of burnsBurns, 2006
- Terahertz Technology in Biology and MedicineIEEE Transactions on Microwave Theory and Techniques, 2004
- Microvascular Assessment of Burn Depth Conversion during Varying Resuscitation ConditionsJournal of Burn Care & Rehabilitation, 2001
- Progressive Burn Injury Documented with Vimentin ImmunostainingJournal of Burn Care & Rehabilitation, 1996
- Pathophysiologic Events Related to Thermal Injury of Skin1990