Direct access transcatheter mitral annuloplasty with a sutureless and adjustable device: preclinical experience
Open Access
- 4 April 2012
- journal article
- research article
- Published by Oxford University Press (OUP) in European Journal of Cardio-Thoracic Surgery
- Vol. 42 (3) , 524-529
Abstract
The aim of the study was to evaluate the technical feasibility and performance of a transcatheter mitral annuloplasty system. Adult swines (n = 15) underwent left thoracotomy through the 4th–5th intercostal space. A transcatheter device (CardioBand, Valtech-Cardio Ltd) was introduced through an 18F sheath through the left atrium and attached to the annulus between the posterior and anterior commissures using echocardiographic and fluoroscopic guidance, on the beating heart. The sutureless device was implanted using a steerable delivery system to deploy sequential fixation elements. Following implantation, the device length was adjusted on the beating heart to reduce the intercommissural and septolateral dimension, under echocardiographic guidance. Finally, the flexible adjustment tool was withdrawn from the working sheath and the atrial purse-string closed. All but five animals were sacrificed acutely by intent, while the others were sacrificed at 90 days. All animals survived the acute implant. One animal died at the third post-operative day due to bleeding. The annuloplasty system was successfully implanted in all animals. A mean of 12 ± 3 fixation elements were deployed. The band length was reduced up to 20% after implantation in each animal. At necropsy, the location of the implant was within a few millimetres of the annulus (3.5 ± 4 mm). In three animals, fixation elements were implanted inadvertently in the leaflets, but no coronary lesions were observed. All animals survived the acute implant. One animal died on the third post-operative day due to bleeding. In the four long-term survivors, the implanted annuloplasty device showed satisfactory healing and no ring dehiscence. Transcatheter minimally invasive, beating-heart implantation of an adjustable annuloplasty band is feasible in the animal model. This approach may be an alternative to open surgical procedures in high-risk patients.Keywords
This publication has 20 references indexed in Scilit:
- Percutaneous Repair or Surgery for Mitral RegurgitationThe New England journal of medicine, 2011
- Transapical repair of mitral valve paravalvular leakage using 3‐D transesophageal guidanceCatheterization and Cardiovascular Interventions, 2010
- On the Use of CartoSound for Left Atrial NavigationJournal of Cardiovascular Electrophysiology, 2010
- The Implantable Cardioverter‐Defibrillator Lead: Principles, Progress, and PromisesPacing and Clinical Electrophysiology, 2009
- Do small (6.6 Fr.) active and passive fixation defibrillation leads perform as well as larger sized leads? A multi-centre analysisEP Europace, 2007
- What are the characteristics of patients with severe, symptomatic, mitral regurgitation who are denied surgery?European Heart Journal, 2007
- Intracardiac Echocardiography for Percutaneous Mitral Valve Repair in a Swine ModelJournal of the American Society of Echocardiography, 2006
- Recurrence of Mitral Valve Regurgitation After Mitral Valve Repair in Degenerative Valve DiseaseCirculation, 2003
- The double-orifice technique in mitral valve repair: A simple solution for complex problemsThe Journal of Thoracic and Cardiovascular Surgery, 2001
- Durability of mitral valve repair for degenerative diseaseThe Journal of Thoracic and Cardiovascular Surgery, 1998