Angular dependence of detector responses in high-energy brachytherapy
Open Access
- 2 June 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (11) , 115020
Abstract
Objective. To characterize three small volume ionization chambers and a diamond detector concerning their angular dependent response under Ir-192 irradiation. Approach. Monte Carlo (MC) simulations as well as experimental measurements of the angular dependent response were performed at 10 cm distance in air. Further simulations under TG-43 conditions were performed at distances of 1, 3, 5 and 10 cm in water. The detectors were placed in relation to their reference points specified by the manufacturer and rotated relative to the beam axis. Results for the different setups were compared to investigate the influence of water and air as well as a distance dependence of the angular response. Detector specific properties such as dead volumes of ionization chambers or asymmetries were taken into account. Main results. The MC simulations in air could be experimentally verified. The detectors showed an angular dependence < 0.5 % for rotations ⩽ 20 ∘ , ⩽ 35 ∘ , ⩽ 40 ∘ and ⩽ 50 ∘ at distances of 1, 3, 5 and 10 cm in water, respectively. In air at 10 cm distance, the angular dependence exceeded 0.5% for rotations of 20∘ for the microDiamond and for 50∘ for the Semiflex and PinPoint. Here, the angular dependence of the A26 was < 0.5 % at all angles. The slope of the angular dependence increased with rotation from frontal alignment for all detectors. Variations of the detector angular dependent response caused by volume averaging were largest close to the source, i.e. up to 8% at 1 cm distance. High-Z materials in the vicinity of the sensitive volumes were an additional source of angular response variations amounting up to 4% independent of the distance. Significance. To the best of the authors’ knowledge, this work is the first investigation of perturbation factors, also differential in energy, leading to angular dependent detector response in high-energy brachytherapy dosimetry. The results could be useful to estimate the angular dependent response when measuring complex dose rate distributions, for example multiple source positions or in a heterogeneous environment.Keywords
This publication has 52 references indexed in Scilit:
- Beam quality correction factors for dose measurements around 192Ir brachytherapy sourcesJournal of Applied Clinical Medical Physics, 2024
- Update of the CLRP Monte Carlo TG‐43 parameter database for high‐energy brachytherapy sourcesMedical Physics, 2023
- Effect of angular dependence for small‐field dosimetry using seven different detectorsMedical Physics, 2023
- Dosimetry in 1.5 T MR‐Linacs: Monte Carlo determination of magnetic field correction factors and investigation of the air gap effectMedical Physics, 2022
- Cema‐based formalism for the determination of absorbed dose for high‐energy photon beamsMedical Physics, 2021
- The role of the construction and sensitive volume of compact ionization chambers on the magnetic field‐dependent dose responseMedical Physics, 2021
- Suitability of the microDiamond detector for experimental determination of the anisotropy function of High Dose Rate 192Ir brachytherapy sourcesMedical Physics, 2020
- A Monte Carlo study on the PTW 60019 microDiamond detectorMedical Physics, 2019
- Suitability of microDiamond detectors for the determination of absorbed dose to water around high‐dose‐rate 192Ir brachytherapy sourcesMedical Physics, 2017
- RECORDS: improved Reporting of montE CarlO RaDiation transport Studies: Report of the AAPM Research Committee Task Group 268Medical Physics, 2017