Preoperative, Computer Simulation-Based, Three-Dimensional Corrective Osteotomy for Cubitus Varus Deformity with Use of a Custom-Designed Surgical Device
Authors:
- Kunihiro OkaKOKunihiro Oka
- Junichi MiyakeJMJunichi Miyake
- Toshiyuki KataokaTKToshiyuki Kataoka
- Hisao MoritomoHMHisao Moritomo
- Tsuyoshi MuraseTMTsuyoshi Murase
- 20 November 2013
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 95 (22) , e173
Abstract
De system involving a custom-made surgical template designed on the basis of a three-dimensional computer simulation incorporating computed tomography (CT) data. We aimed to investigate the feasibility of this novel technique for correcting cubitus varus deformity. Methods: Thirty consecutive patients (twenty-three males and seven females) with a cubitus varus deformity resulting from the malunion of a distal humeral supracondylar fracture were included in this study. Between October 2003 and May 2011, the patients underwent a three-dimensional corrective osteotomy with use of a custom-made surgical template. The patients were then followed for a minimum of twelve months. We evaluated radiographic parameters, including the humerus-elbow-wrist angle and tilting angle, as well as the ranges of motion of the elbow and shoulder at the time of the most recent follow-up. An overall clinical evaluation was performed. Results: Bone union was achieved at a mean of four months after surgery. The mean humerus-elbow-wrist angle and tilting angle on the affected side improved significantly from 18.2° (varus) and 25.0°, respectively, before surgery, to 5.8° (valgus) and 38.0°, respectively, after surgery. Hyperextension of the elbow and internal rotation of the shoulder were normalized in all patients. Early plate breakage was observed in one patient. One patient had mild recurrence of varus deformity. Twenty-seven patients had an excellent result, three had a good result, and none had a poor result. Conclusions: Three-dimensional corrective osteotomy with the use of a custom-made surgical template that is designed and produced on the basis of computer simulation is a feasible and useful treatment option for cubitus varus deformity. Level of Evidence: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence....Keywords
This publication has 38 references indexed in Scilit:
- Three-dimensional analysis of cubitus varus deformity after supracondylar fractures of the humerusJournal of Shoulder and Elbow Surgery, 2011
- Application of computer-aided design osteotomy template for treatment of cubitus varus deformity in teenagers: A pilot studyJournal of Shoulder and Elbow Surgery, 2011
- Corrective dome osteotomy for cubitus varus and valgus in adultsJournal of Shoulder and Elbow Surgery, 2009
- Oblique Closing Wedge Osteotomy and Lateral Plating for Cubitus Varus in AdultsClinical Orthopaedics and Related Research, 2008
- Lateral closed wedge osteotomy for cubitus varus deformityIndian Journal of Orthopaedics, 2008
- Tardy ulnar nerve palsy in cubitus varus deformity associated with ulnar nerve dislocation in adultsJournal of Shoulder and Elbow Surgery, 2006
- Three-dimensional corrective osteotomy for cubitus varus in adultsJournal of Shoulder and Elbow Surgery, 2003
- Tardy ulnar nerve palsy caused by cubitus varus deformityThe Journal of Hand Surgery, 1995
- Internal rotation deformity and tardy ulnar nerve palsy after supracondylar humeral fractureJournal of Shoulder and Elbow Surgery, 1995
- Three-dimensional corrective osteotomy for treatment of cubitus varus after supracondylar fracture of the humerus in childrenJournal of Shoulder and Elbow Surgery, 1995