Handheld Device Adapted to Smartphone Cameras for the Measurement of Sodium Ion Concentrations at Saliva-Relevant Levels via Fluorescence
Open Access
- 2 June 2015
- journal article
- research article
- Published by MDPI AG in Bioengineering
- Vol. 2 (2) , 122-138
Abstract
The use of saliva sampling as a minimally-invasive means for drug testing and monitoring physiology is a subject of great interest to researchers and clinicians. This study describes a new optical method based on non-axially symmetric focusing of light using an oblate spheroid sample chamber. The device is simple, lightweight, low cost and is easily attached to several different brands/models of smartphones (Apple, Samsung, HTC and Nokia) for the measurement of sodium ion levels at physiologically-relevant saliva concentrations. The sample and fluorescent reagent solutions are placed in a specially-designed, lightweight device that excludes ambient light and concentrates 470-nm excitation light, from a low-power photodiode, within the sample through non-axially-symmetric refraction. The study found that smartphone cameras and post-image processing quantitated sodium ion concentration in water over the range of 0.5–10 mM, yielding best-fit regressions of the data that agree well with a data regression of microplate luminometer results. The data suggest that fluorescence can be used for the measurement of salivary sodium ion concentrations in low-resource or point-of-care settings. With further fluorescent assay testing, the device may find application in a variety of enzymatic or chemical assays.Keywords
This publication has 21 references indexed in Scilit:
- Comparison of scattering diagrams of large non-spherical particles calculated by VCRM and MLFMAJournal of Quantitative Spectroscopy and Radiative Transfer, 2015
- Recent advances in cortisol sensing technologies for point-of-care applicationBiosensors and Bioelectronics, 2014
- Application of Newton’s Zero Order Caustic for Analysis and Measurement: Part-I AbsorbanceInternational Research Journal of Pure and Applied Chemistry, 2014
- Saliva-Based Biosensors: Noninvasive Monitoring Tool for Clinical DiagnosticsBioMed Research International, 2014
- Label-free biodetection using a smartphoneLab on a Chip, 2013
- Simulated Biological Fluids with Possible Application in Dissolution TestingDissolution Technologies, 2011
- Use of Colorimetric Test Strips for Monitoring the Effect of Hemodialysis on Salivary Nitrite and Uric Acid in Patients with End-Stage Renal Disease: A Proof of PrincipleClinical Chemistry, 2008
- Axially symmetric focusing as a cuspoid diffraction catastrophe: Scalar and vector cases and comparison with the theory of MiePhysical Review B, 2006
- Sodium Green as a Potential Probe for Intracellular Sodium Imaging Based on Fluorescence LifetimeAnalytical Biochemistry, 1997
- Hyperbolic umbilic diffraction catastrophe and rainbow scattering from spheroidal dropsNature, 1984