Wearable Cuff-Less Blood Pressure Estimation at Home via Pulse Transit Time
- 3 June 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. 25 (6) , 1926-1937
Abstract
Objective: We developed a wearable watch-based device to provide noninvasive, cuff-less blood pressure (BP) estimation in an at-home setting. Methods: The watch measures single-lead electrocardiogram (ECG), tri-axial seismocardiogram (SCG), and multi-wavelength photoplethysmogram (PPG) signals to compute the pulse transit time (PTT), allowing for BP estimation. We sent our custom watch device and an oscillometric BP cuff home with 21 healthy subjects, and captured the natural variability in BP over the course of a 24-hour period. Results: After calibration, our Pearson correlation coefficient (PCC) of 0.69 and root-mean-square-error (RMSE) of 2.72 mmHg suggest that noninvasive PTT measurements correlate with around-the-clock BP. Using a novel two-point calibration method, we achieved a RMSE of 3.86 mmHg. We further demonstrated the potential of a semi-globalized adaptive model to reduce calibration requirements. Conclusion: This is, to the best of our knowledge, the first time that BP has been comprehensively estimated noninvasively using PTT in an at-home setting. We showed a more convenient method for obtaining ambulatory BP than through the use of the standard oscillometric cuff. We presented new calibration methods for BP estimation using fewer calibration points that are more practical for a real-world scenario. Significance: A custom watch (SeismoWatch) capable of taking multiple BP measurements enables reliable remote monitoring of daily BP and paves the way towards convenient hypertension screening and management, which can potentially reduce hospitalizations and improve quality of life.Keywords
Funding Information
- National Institutes of Health
- National Institute of Biomedical Imaging and Bioengineering (1U01EB018818-01)
- National Institute of Biomedical Imaging and Bioengineering Point-of-Care Technologies Research Network
- Atlanta Center for Microsystems-Engineered Point-of-Care Technologies (U54EB027690)
This publication has 27 references indexed in Scilit:
- Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a referencePhysiological Reports, 2016
- Screening for High Blood Pressure in Adults: U.S. Preventive Services Task Force Recommendation StatementAnnals of Internal Medicine, 2015
- [Mobile Health: IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices].2015
- Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and PracticeIEEE Transactions on Biomedical Engineering, 2015
- A Wearable Patch to Enable Long-Term Monitoring of Environmental, Activity and Hemodynamics VariablesIEEE Transactions on Biomedical Circuits and Systems, 2015
- Cost–Benefit Analysis of Home Blood Pressure Monitoring in Hypertension Diagnosis and TreatmentHypertension, 2014
- Progression to hypertension in non-hypertensive children following renal transplantationNephrology Dialysis Transplantation, 2012
- An independent relationship between muscle sympathetic nerve activity and pulse wave velocity in normal humansJournal Of Hypertension, 2010
- Evaluating the Lower-Body Electromyogram Signal Acquired From the Feet As a Noise Reference for Standing Ballistocardiogram MeasurementsIEEE Transactions on Information Technology in Biomedicine, 2010
- Heart rate variability: Origins, methods, and interpretive caveatsPsychophysiology, 1997