Positive corona discharge ion source with IMS/MS to detect impurities in high purity Nitrogen
- 21 July 2011
- journal article
- research article
- Published by EDP Sciences in The European Physical Journal Applied Physics
- Vol. 55 (1) , 13808
Abstract
We have applied the Ion Mobility Spectrometry/Mass Spectrometry (IMS/MS) and the Atmospheric Pressure Chemical Ionisation/Mass Spectrometry (APCI/MS) techniques to study the formation of the ions in the positive corona discharge (CD) in highly purified nitrogen with impurities at 100 ppt level. The main products observed were H3O+(H2O) n ions (reduced ion mobility of 2.15 cm2 V-1 s-1). Additionally, we have observed ions with reduced mobilities 2.42 cm2 V-1 s-1 and 2.30 cm2 V-1 s-1. The intensity of these ions was increasing with the increasing discharge current. We associated these peaks with NH4 + and NO+(H2O) n . The formation of these ions results from trace amounts of O2 and NH3 in nitrogen. The time evolution of the ions in corona discharge has been studied using the APCI/MS technique in the time windows ranging from 100 μs to 2 ms. The present work indicates the ability of the IMS technique equipped with CD ions source to detect impurities below 100 ppt level.Keywords
This publication has 19 references indexed in Scilit:
- Detection of explosives by positive corona discharge ion mobility spectrometryJournal of Hazardous Materials, 2010
- Improvement in Analysis of O2 in N2 by Using Ar Drift Gas in an Ion Mobility SpectrometerAnalytical Chemistry, 2002
- Use of Ion Mobility Spectrometry To Determine Low Levels of Impurities in GasesAnalytical Chemistry, 2001
- Design and optimization of a corona discharge ionization source for ion mobility spectrometryReview of Scientific Instruments, 2000
- Generation of [H3O]+.(H2O)n clusters by positive corona discharge in airRapid Communications in Mass Spectrometry, 1997
- Abatement of volatile organic compounds by corona discharge. A study of the reactivity of trichloroethylene under atmospheric pressure ionization conditionsRapid Communications in Mass Spectrometry, 1997
- Quantitative analysis of contaminants in ultrapure gases at the parts-per-trillion level using atmospheric pressure ionization mass spectroscopyJournal of Vacuum Science & Technology A, 1993
- Identification of positive reactant ions observed for nitrogen carrier gas in plasma chromatograph mobility studiesAnalytical Chemistry, 1975
- Solvation of the hydrogen ion by water molecules in the gas phase. Heats and entropies of solvation of individual reactions. H+(H2O)n-1 + H2O .fwdarw. H+(H2O)nJournal of the American Chemical Society, 1967
- Mass-Spectrometric Studies of Corona Discharges in Air at Atmospheric PressuresThe Journal of Chemical Physics, 1966