The use of two different buffers for unequivocal molecular weight determination in liquid chromatography/thermospray mass spectrometry
Author:
- 1 June 1989
- journal article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 3 (6) , 183-185
Abstract
High‐performance liquid chromatography (HPLC) in combination with thermospray mass spectrometry (LC/TSP‐MS) is an important method for molecular weight determination of unknown compounds. In addition to the well‐known ammonium acetate buffer (NH 4 OAc) the use of methylammonium acetate (CH 3 NH 3 OAc) in the positive‐ and ammonium formate (NH 4 OFo) in the negative‐ion mode is recommended. Performing a second run with these buffers, even from spectra with only one single quasimolecular ion, an unequivocal molecular weight can be obtained. For correct assignment, the shift in mass of the corresponding quasimolecular ions in each mode is used. Finally, the use of a second buffer greatly facilitates the recognition of superimposed HPLC peaks.Keywords