Continuous Glucose Monitoring System in a Rural Intensive Care Unit: A Pilot Study Evaluating Accuracy and Acceptance
Authors:
Open Access
- 1 May 2010
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 4 (3) , 636-644
Abstract
Glucose management in an intensive care unit (ICU) is labor-intensive. A continuous glucose monitoring system (CGMS) has the potential to improve efficiency and safety in this setting. The goal of this study was to determine if the Medtronic Guardian® REAL-Time CGMS was accurate and tolerated by patients in a rural hospital ICU unit. Differences between individual finger stick blood glucose (FSBG) and CGMS values were compared to American Diabetes Association (ADA) and International Organization for Standardization (ISO) standards. Continuous glucose monitoring system accuracy was evaluated over four ranges: 200 mg/dl. Other accuracy measures [mean absolute deviation (MAD), mean absolute relative difference (MARD), and coefficient of linear regression of CGMS on FSBG] were calculated. Nursing staff and patients were surveyed regarding use of the CGMS in the ICU. Twenty-nine participants had 320 FSBG and corresponding CGMS readings. Sixty-two percent of participants were admitted with diabetic ketoacidosis (DKA). Two hundred and thirteen (66.6%) were accurate within the ISO standard, whereas only 70 out of 320 (21.9%) were within the 5% ADA standard. The CGMS was most accurate in euglycemia. Technical difficulties, such as adequate time for “wetting” and calibration of electrodes, arose with the sensors. The MAD was 28.3 mg/dl, the MRAD was 17.4%, and the linear regression coefficient of CGMS on FSBG was 0.834 (p < 0.001). The CGMS is well tolerated by ICU patients but, at present, is not sufficiently accurate to be used for therapeutic decisions in the acute setting, particularly in patients with diabetic ketoacidosis. There is a need to find resolution to the technical issues regarding electrode “wetting” and calibration if CGMS use in the ICU setting is to provide an effective means of diabetes care and management.Keywords
This publication has 29 references indexed in Scilit:
- Standards of Medical Care in Diabetes—2010Diabetes Care, 2010
- Intensive versus Conventional Glucose Control in Critically Ill PatientsNew England Journal of Medicine, 2009
- Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose MonitorsDiabetes Care, 2008
- A replicable method for blood glucose control in critically Ill patientsCritical Care Medicine, 2008
- Continuous Glucose Monitoring in Children with Type 1 DiabetesThe Journal of Pediatrics, 2007
- The role of continuous glucose monitoring in clinical decision‐making in diabetes in pregnancyAustralian and New Zealand Journal of Obstetrics and Gynaecology, 2007
- Real-Time Continuous Glucose Monitoring in Pediatric Patients During and After Cardiac SurgeryPediatrics, 2006
- Evaluation of Factors Affecting CGMS CalibrationDiabetes Technology & Therapeutics, 2006
- Association Between Hyperglycemia and Increased Hospital Mortality in a Heterogeneous Population of Critically Ill PatientsMayo Clinic Proceedings, 2003
- Continuous Subcutaneous Glucose Monitoring in Children With Type 1 DiabetesPediatrics, 2001