Whole-Brain CT Perfusion to Quantify Acute Ischemic Penumbra and Core

Authors:

Abstract
Caution must be applied in the interpretation of limited-coverage CT perfusion, with the accuracy of the penumbra and core measurement called into question when section coverage is less than 80 mm. Purpose To validate the use of perfusion computed tomography (CT) with whole-brain coverage to measure the ischemic penumbra and core and to compare its performance to that of limited-coverage perfusion CT. Materials and Methods Institutional ethics committee approval and informed consent were obtained. Patients (n = 296) who underwent 320-detector CT perfusion within 6 hours of the onset of ischemic stroke were studied. First, the ischemic volume at CT perfusion was compared with the penumbra and core reference values at magnetic resonance (MR) imaging to derive CT perfusion penumbra and core thresholds. Second, the thresholds were tested in a different group of patients to predict the final infarction at diffusion-weighted imaging 24 hours after CT perfusion. Third, the change in ischemic volume delineated by the optimal penumbra and core threshold was determined as the brain coverage was gradually reduced from 160 mm to 20 mm. The Wilcoxon signed-rank test, concordance correlation coefficient (CCC), an...