A Steam Sterilizable Plastic-Encapsulated Wireless Sensor Module
- 12 February 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Components, Packaging and Manufacturing Technology
- Vol. 9 (4) , 770-778
Abstract
This paper presents a plastic-encapsulated wireless sensor module (WSM), using commercial-off-the-shelf (COTS) components, that has been demonstrated to be reliable for 180 cycles in a medical steam sterilizer with an environment involving high-rate temperature, pressure and humidity cycles up to 134°C, 3 bar, 100% RH. The WSM continuously transmits temperature and pressure data from the sterilizer, allowing the quality of sterilization to be verified. Twelve WSMs were assembled and encapsulated with substrate material and encapsulant type as experiment variables. These samples were then stress-tested over multiple sterilization cycles, with continuous wireless data collection, until failure. Scanning Acoustic Microscope (SAM) analysis was carried out before and after the reliability testing in the steam sterilizer. The results demonstrate that, with appropriate choice of materials, the WSM can survive for 180 full steam sterilization cycles without any electrical failure of the components or any deterioration in the package structure except for delamination between the substrate and the encapsulant that did not cause electrical failure. The overall results demonstrate the reliability of a plastic-encapsulated module in an exceptionally harsh use environment, the ability to steam sterilize a plastic-encapsulated module for medical applications, and the potential for using a plastic-encapsulated WSM for continuous, high-resolution sterilizer parameter monitoring as a replacement for the currently used color-change chemical and biological indicators. This is believed to be the first plastic-encapsulated WSM for monitoring of steam sterilizer parameters and the first plastic-encapsulated module of any sort which is capable of surviving multiple steam sterilization cycles..Keywords
Funding Information
- Higher Education Authority of Ireland of the Program for Research in Third Level Institutions, Cycle IV Project NEMBES
This publication has 16 references indexed in Scilit:
- An electronic encapsulated monitoring system for a vascular graft bioreactor2015
- Reliability of plastic-encapsulated electronic components in supersaturated steam environmentsIEEE Transactions on Components, Packaging and Manufacturing Technology, 2015
- Analysis of Dynamic Moisture Diffusion in Packaging Materials During Steam SterilizationIEEE Transactions on Components, Packaging and Manufacturing Technology, 2015
- Reliability analysis of plastic encapsulated coin batteries under harsh environmental conditions2012
- Multi-physics modeling of LED-based luminaires under temperature and humidity environment2012
- Reliability testing of electronic packages in harsh environments2010
- Measurement of Moisture-Induced Packaging Stress With Piezoresistive SensorsIEEE Transactions on Advanced Packaging, 2007
- Performance of various steam sterilization indicators under optimum and sub-optimum exposure conditionsAmerican Journal of Infection Control, 2005
- Integrated vapor pressure, hygroswelling, and thermo-mechanical stress modeling of QFN package during reflow with interfacial fracture mechanics analysisMicroelectronics Reliability, 2004
- Biological indicators for steam sterilization: characterization of a rapid biological indicator utilizingBacillus stearothermophilusspore-associated alpha-glucosidase enzymeJournal of Applied Microbiology, 1998