Two-Step Validation of a New Wireless Inertial Sensor System: Application in the Squat Motion
Open Access
- 9 June 2022
- journal article
- research article
- Published by MDPI AG in Technologies
- Vol. 10 (3) , 72
Abstract
The use of Inertial Measurement Units (IMUs) can provide embedded motion data to improve clinical application. The objective of this study was to validate a newly designed IMU system. The validation is provided through two main methods, a classical sensor validation achieved on a six-degrees-of-freedom hexapod platform with controlled linear and rotation motions and a functional validation on subjects performing squats with segmental angle measurement. The kinematics of the sensors were measured by using an optoelectronic reference system (VICON) and then compared to the orientation and raw data of the IMUs. Bland–Altman plots and Lin’s concordance correlation coefficient were computed to assess the kinematic parameter errors between the IMUs and VICON system. The results showed suitable precision of the IMU system for linear, rotation and squat motions.Keywords
Funding Information
- French Ministry of Higher Education, Research and Innovation (4.7.360)
- National Centre for Scientific Research (CNRS) (4.7.360)
- NOMADe Project of the EU Interreg Program (4.7.360)
This publication has 47 references indexed in Scilit:
- Inertial Measures of Motion for Clinical Biomechanics: Comparative Assessment of Accuracy under Controlled Conditions - Effect of VelocityPLOS ONE, 2013
- Avancées thérapeutiques de la lombalgie chronique : revue systématique de la littérature des 5 dernières annéesAnnals of Physical and Rehabilitation Medicine, 2012
- The difference between reliability and agreementJournal of Clinical Epidemiology, 2011
- A Consensus Approach Toward the Standardization of Back Pain Definitions for Use in Prevalence StudiesSpine, 2008
- A systematic review of low back pain cost of illness studies in the United States and internationallyThe Spine Journal, 2008
- Adjustments to McConville et al. and Young et al. body segment inertial parametersJournal of Biomechanics, 2007
- A 3D Generic Inverse Dynamic Method using Wrench Notation and Quaternion AlgebraComputer Methods in Biomechanics and Biomedical Engineering, 2004
- Back painRheumatology, 2002
- Comparison of Lumbar Range of Motion Using Three Measurement Devices in Patients With Chronic Low Back PainSpine, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986