Direct clinical comparison of three commercial BK polyomavirus DNA assays in plasma specimens: Alinity m BKV, cobas BKV, and Panther Fusion BKV

Authors:

Abstract
BackgroundReactivation of BK polyomavirus (BKPyV) in kidney transplant recipients can lead to BKPyV-associated nephropathy (BKPyVAN), a major cause of allograft dysfunction. Quantitative nucleic acid amplification tests (QNATs) enable monitoring, yet inter-assay variability may complicate interpretation, despite standardization to the WHO International Standard.ObjectiveThis study compares the qualitative agreement, quantitative correlation, and concordance at clinically relevant thresholds across three automated assays: Alinity m BKV, cobas BKV, and Panther Fusion BKV.MethodsA total of 160 remnant EDTA plasma specimens from kidney transplant recipients were tested across the three platforms, yielded valid results on Alinity m BKV and cobas BKV (n = 160 each) and Panther Fusion BKV (n = 152). Pairwise analysis included qualitative agreement, Cohen’s κ, Deming regression, Bland-Altman bias and agreement at clinically relevant thresholds of ≤ 4 and > 4 log10 IU/mL.ResultsOverall qualitative agreement ranged from 92.1% to 96.1% with κ indicating substantial to near-perfect agreement. Quantitative correlations were strong (r ≥ 0.929), with minimal mean bias (-0.33 to -0.08 log10 IU/mL). Agreement was higher at ≤ 4 log10 IU/mL interval and moderate at > 4 log10 IU/mL. Discordances clustered near the quantitation limit of the assays.ConclusionThe three automated assays demonstrated high qualitative agreement and closely aligned quantitative result, with minor systematic differences near the quantitation limit. Given that the commercial assays report in IU/mL while guideline thresholds have historically been expressed in copies/mL, clinical decisions are best supported by assay-consistent longitudinal trends rather than cross-platform threshold comparisons.