Evaluation of electric neurostimulation to confirm correct placement of lumbosacral epidural injections in dogs
Authors:
- Fernando L. Garcia-PereiraFGFernando L. Garcia-Pereira
- Joe HauptmanJHJoe Hauptman
- Andre C. ShihASAndre C. Shih
- Sharon E. LairdSLSharon E. Laird
- Anthony PeaseAPAnthony Pease
- 1 February 2010
- journal article
- clinical trial
- Published by American Veterinary Medical Association (AVMA) in American Journal of Veterinary Research
- Vol. 71 (2) , 157-160
Abstract
Objective—To determine the minimal electric threshold (MET) of neurostimulation in and out of the lumbosacral epidural space necessary to cause muscle contraction of the hind limb or tail, determine an MET cutoff value that indicates epidural needle placement, and compare predictability of epidural needle placement attained by use of neurostimulation versus the standard technique that uses loss of resistance in dogs. Animals—96 healthy Beagles. Procedures—Dogs received nonionic contrast medium (90 mg/kg) either in or out of the epidural space. Correct placement of the needle was evaluated by use of neurostimulation and loss of resistance of injection and confirmed by use of epidurography. Results—With the neurostimulator test, MET was significantly lower in dogs with needle placement in the epidural space (mean ± SEM, 0.30 ± 0.07 mA) than those with needle placement out of the epidural space (1.2 ± 0.13 mA). When an electric current cutoff of ≤ 0.28 mA for the neurostimulator test was used to suggest correct needle placement in the lumbosacral epidural space, sensitivity and specificity were 74% and 93%, respectively. The loss of resistance test had sensitivity of 63% and specificity of 90%. The combination of both tests yielded a sensitivity of 89% and specificity of 83%. Conclusions and Clinical Relevance—Neurostimulation is a useful tool to suggest correct lumbosacral epidural needle placement in dogs.Keywords
This publication has 14 references indexed in Scilit:
- Use of pressure waves to confirm the correct placement of epidural needles in dogsVeterinary Record, 2007
- Innovative Approaches to Neuraxial Blockade in Children: The Introduction of Epidural Nerve Root Stimulation and Ultrasound Guidance for Epidural Catheter PlacementPain Research and Management, 2006
- The Use of Electrical Stimulation to Monitor Epidural Needle Advancement in a Porcine ModelAnesthesia & Analgesia, 2005
- Lateral cervical epidural catheter placement using nerve stimulation for continuous unilateral upper extremity analgesia following a failed continuous peripheral nerve blockActa Anaesthesiologica Scandinavica, 2005
- The threshold current in the intrathecal space to elicit motor response is lower and does not overlap that in the epidural space: a porcine modelCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2004
- Epidural Analgesia in the Dog and CatThe Veterinary Journal, 2001
- Detection of subarachnoid and intravascular epidural catheter placementCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1999
- Confirmation of epidural catheter placement using nerve stimulationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1998
- Combined epidural and peripheral nerve stimulation for relief of painJournal of Neurosurgery, 1982
- RECORDINGS OF, AND AN ANATOMICAL EXPLANATION FOR, FALSE POSITIVE LOSS OF RESISTANCE DURING LUMBER EXTRADURAL ANALGESIABritish Journal of Anaesthesia, 1979