A dual-wavelength method using the BDJ detector and its application to iron concentration measurement
Author:
- 1 January 1999
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 10 (4) , 312-315
Abstract
A new and uncomplicated method for absorption detection is presented that requires no optical instrument such as a spectrophotometer or colorimeter. It makes use of a two-wavelength light source and a BDJ (buried double p-n junction) detector operating as wavelength-sensitive device. The intensity-insensitive photocurrent ratio rdual=I2/I1 is used as the detector response and is related to the absorbing species. A reference curve rdual(C) obtained by calibration serves to determine concentrations of the component. It has been theoretically shown and experimentally verified that the curve rdual(C) becomes linear at low concentrations. Through an investigation in the case of iron concentration measurements, the suggested method has proved to be sensitive and reliable. Keywords: BDJ detector, dual-wavelength method, absorption detection, concentration measurementsKeywords
This publication has 7 references indexed in Scilit:
- A colorimetric method with the use of BDJ detector for seawater pH measurementAnalytica Chimica Acta, 1998
- A seawater pH determination method using a BDJ detectorMeasurement Science and Technology, 1998
- Buried double p-n junction structure using a CMOS process for wavelength detection1997
- Investigation of a buried double p-n junction structure implemented in CMOS technology for wavelength-sensitive detectionInternational Journal of Electronics, 1997
- Colour detection using a buried double p - n junction structure implemented in the CMOS processElectronics Letters, 1996
- Temperature-dependent minority-carrier lifetime measurements of red AlGaAs light emitting diodesJournal of Electronic Materials, 1995
- Wavelength selection strategies to enhance referencing in LED based optical sensorsOptics Communications, 1994