Tensile Fixation Strengths of Absorbable Meniscal Repair Devices as a Function of Hydrolysis Time
Authors:
- 1 March 2001
- journal article
- research article
- Published by SAGE Publications in The American Journal of Sports Medicine
- Vol. 29 (2) , 118-123
Abstract
To determine the effect of hydrolysis time on the fixation strength of absorbable meniscal repair devices, adult bovine menisci were repaired with five devices and a suture. The ultimate tensile strength of the repair was then tested in six specimens immediately or after 6, 12, or 24 weeks of incubation at 37°C in a saline solution containing antibiotics, antimycotics, and protease inhibitors. Immediately after implantation the Bionx Meniscus Arrow had a significantly higher failure strength (57.7 13.8 N) than the Linvatec BioStinger (35.1 6.7 N), the Innovasive Clearfix screw (34.9 13 N), the Surgical Dynamics S D sorb staple (9.4 4.6 N), and the Mitek Meniscal Repair System (polydioxanone) (27.2 6.0 N). However, there was no significant difference between the Bionx Meniscus Arrow and a 2—0 polydioxanone vertical suture (51.6 2.7 N). The polydioxanone-based implants demonstrated a significant decrease in failure strength at 12 and 24 weeks. Similarly, the Surgical Dynamics S D sorb staple lost all fixation strength by 24 weeks. The remaining devices showed no significant loss of failure strength over the 24-week period, suggesting that 24 weeks of hydrolysis does not adversely affect the ultimate holding power of poly L-lactide-based meniscal fixation devices.Keywords
This publication has 19 references indexed in Scilit:
- Pull-Out Strength and Stiffness of Meniscal Repair Using Absorbable Arrows or Ti-Cron Vertical and Horizontal Loop SuturesThe American Journal of Sports Medicine, 1999
- Broken Arrow: A Complication of Meniscal RepairArthroscopy, 1999
- Meniscus RepairThe American Journal of Sports Medicine, 1999
- Failure of a Biodegradable Meniscal ArrowThe American Journal of Sports Medicine, 1999
- Mechanical Comparison of 10 Suture Materials before and after in Vivo IncubationJournal of Surgical Research, 1994
- Accelerated rehabilitation for meniscus repairsArthroscopy, 1994
- The effect of inflammatory synovial fluid on the breaking strength of new “long lasting” absorbable suturesArthroscopy, 1992
- In vivo and in vitro degradatation of monofilament absorbable sutures, PDS® and Maxon®Biomaterials, 1990
- Inflammatory synovial fluid and absorbable suture strengthArthroscopy, 1988
- The microvasculature of the meniscus and its response to injuryThe American Journal of Sports Medicine, 1983