Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality

Authors:

Abstract
OBJECTIVE. The purpose of this study was to evaluate the efficiency of automatic bone removal in dual-energy CT angiography (CTA) of the trunk. SUBJECTS AND METHODS. Nineteen patients underwent dual-energy CTA of the trunk (tube A, 140 kV; tube B, 100 kV). In addition to the dual-energy dataset, an image equivalent to that of a standard 120-kV single-energy examination was generated with both tubes. Automated bone segmentation was performed on both datasets, and the results were analyzed. The time required for and subjective image quality of the maximum intensity projections (MIPs) generated were evaluated. RESULTS. Errors in bone segmentation were found for 1.5% of bones on dual-energy images and 12.4% of bones on single-energy images (p < 0.01). The most important differences were found in the rib cage, sternum, and pelvis. The times required for postprocessing of MIPs were similar for the dual-energy (113.5 seconds) and single-energy (106.8 seconds) techniques. The subjective image quality of the arter...