Balloon assisted Woven endobridge deployment (BAWD): A safety and efficacy study

Authors:

Abstract
Background: Balloon-assisted deployment/remodelling is a proven adjunctive technique for coil embolization of intracranial aneurysms, and it may be a helpful adjunct in delivering the Woven EndoBridge (WEB) device. Objective: To evaluate the safety, efficacy and feasibility of balloon-assisted WEB deployment in both ruptured and unruptured intracranial aneurysms in both typical and atypical locations. Methods: Patients who underwent treatment of ruptured and unruptured intracranial aneurysms with the BAWD technique were retrospectively identified from a prospectively maintained database at two neurointerventional centres. Patient demographics, aneurysm characteristics, technical procedure details, clinical and imaging outcomes were reviewed. Results: Thirty-three aneurysms (23 women) were identified with a median age of 58 years. There were 15 (45.5%) ruptured aneurysms, 25 (64.3%) in the anterior circulation and 12 (36.4%) aneurysms having an atypical location for WEB treatment. The average aneurysm size was 6.8 mm (greatest dimension), 4.6 mm (height) and 4.5 mm (width), and 25 (75.8%) aneurysms had a wide neck morphology. One patient died (3.0%) secondary to a procedure-related complication, and there was no procedure-related permanent morbidity. Complete and adequate aneurysm occlusion on mid-term follow-up DSA was 85.2% and 92%, respectively. Conclusion: Balloon-assisted WEB deployment appears to be a safe and effective technique that may increase the utility of the WEB device. Further prospective studies on BAWD should be considered.

This publication has 18 references indexed in Scilit: