High resolution extreme ultraviolet spectrometer for an electron beam ion trap
Authors:
- 1 August 2011
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 82 (8) , 083103
Abstract
An extreme ultraviolet spectrometer has been developed for spectroscopic studies of highly charged ions with an electron beam ion trap. It has a slit-less configuration with a spherical varied-line-spacing grating that provides a flat focal plane for grazing incidence light. Alternative use of two different gratings enables us to cover the wavelength range 1–25 nm. Test observations with the Tokyo electron beam ion trap demonstrate the high performance of the present spectrometer such as a resolving power of above 1000.Keywords
This publication has 27 references indexed in Scilit:
- Spectroscopy of highly charged ions in Solar and astrophysical plasmasJournal of Physics: Conference Series, 2009
- Compact soft x-ray spectrometer for plasma diagnostics at the Heidelberg Electron Beam Ion TrapReview of Scientific Instruments, 2007
- EUV Sources for Lithography2006
- A high efficiency ultrahigh vacuum compatible flat field spectrometer for extreme ultraviolet wavelengthsReview of Scientific Instruments, 2005
- Spectroscopic Diagnostics for ITERJournal of Plasma and Fusion Research, 2003
- EUV Spectroscopy of Highly Charged Tungsten Ions Relevant to Hot PlasmasPhysica Scripta, 2001
- Grazing-incidence measurements of L-shell line emission from highly charged Fe in the soft x-ray regionReview of Scientific Instruments, 1999
- Spectroscopy in the extreme ultraviolet on an electron beam ion trapReview of Scientific Instruments, 1999
- Measurement of electron-impact–excitation cross sections for very highly charged ionsPhysical Review Letters, 1988
- Mechanically ruled aberration-corrected concave gratings for a flat-field grazing-incidence spectrographApplied Optics, 1983