Investigation of transverse oscillation method
- 5 June 2006
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 53 (5) , 959-971
Abstract
Conventional ultrasound scanners can display only the axial component of the blood velocity vector, which is a significant limitation when vessels nearly parallel to the skin surface are scanned. The transverse oscillation (TO) method overcomes this limitation by introducing a TO and an axial oscillation in the pulse echo field. The theory behind the creation of the double oscillation pulse echo field is explained as well as the theory behind the estimation of the vector velocity. A parameter study of the method is performed, using the ultrasound simulation program Field II. A virtual linear-array transducer with center frequency 7 MHz and 128 active elements is created, and a virtual blood vessel of radius 6.4 mm is simulated. The performance of the TO method is found around an initial point in the parameter space. The parameters varied are: flow angle, transmit focus depth, receive apodization, pulse length, transverse wave length, number of emissions, signal-to-noise ratio (SNR), and type of echo-canceling filter used. Using an experimental scanner, the performance of the TO method is evaluated. An experimental flowrig is used to create laminar parabolic flow in a blood mimicking fluid, and the fluid is scanned under different flow-to-beam angles. The relative standard deviation on the transverse velocity estimate is found to be less than 10% for all angles between 50deg and 90deg. Furthermore, the TO method is evaluated in the flowrig using pulsatile flow, which resembles the flow in the femoral artery. The estimated volume flow as a function of time is compared to the volume flow derived from a conventional axial method at a flow-to-beam angle of 60deg. It is found that the method is highly sensitive to the angle between the flow and the beam direction. Also, the choice of echo canceling filter affects the performance significantlyKeywords
This publication has 13 references indexed in Scilit:
- Directional velocity estimation using focusing along the flow direction. I: theory and simulationIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2003
- Measurement of the complete (3D) velocity vector of blood flows2003
- Experimental investigation of transverse flow estimation using transverse oscillation2003
- A new estimator for vector velocity estimation [medical ultrasonics]IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2001
- A new method for estimation of velocity vectorsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Validation of a New Blood-Mimicking Fluid for Use in Doppler Flow Test ObjectsUltrasound in Medicine & Biology, 1998
- Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992
- Angle Independent Ultrasonic Detection of Blood FlowIEEE Transactions on Biomedical Engineering, 1987
- Ultrasound Doppler Probing of Flows Transverse with Respect to Beam AxisIEEE Transactions on Biomedical Engineering, 1987
- Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique1985