Enhanced mass resolution method development, validation and assay application to support preclinical studies of a new drug candidate

Authors:

Abstract
A very highly sensitive and highly selective liquid chromatographic/tandem mass spectrometric (LC/MS/MS) method was developed to evaluate and quantify a new drug candidate in different biological matrices. Following a simple plasma protein precipitation using acetonitrile, the post‐treatment samples were analyzed on a C 18 column interfaced with a new generation of triple‐quadrupole mass spectrometer. The recently introduced triple‐quadrupole mass spectrometer, the TSQ Quantum Ultra, with enhanced mass‐resolution capability, demonstrated improved sensitivity (0.05 ng/mL), coupled with suitable accuracy and precision, over a broad linear dynamic range (0.05–1000 ng/mL). A comparison of the assay performance data (dynamic range, calibration curve equation, precision and accuracy) of the enhanced resolution method against a unit resolution method under optimized conditions showed the performance improvement of the enhanced mass resolution method for bioanalytical high‐throughput applications. The enhanced mass resolution method herein described was successfully applied to the evaluation of the pharmacokinetic profile of a new drug candidate in rat, rabbit and dog plasma samples. Copyright © 2006 John Wiley & Sons, Ltd.

This publication has 10 references indexed in Scilit: