Intraocular pressure according to different types of tonometry (non-contact and Goldmann applanation) in patients with different degrees of bilateral tearing
Authors:
Open Access
- 16 September 2019
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 14 (9) , e0222652
Abstract
To investigate intraocular pressure (IOP) readings by non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT) for patients with different degrees of bilateral tearing. In this study, we reviewed the medical charts of patients complaining of different degrees of bilateral tearing. The tear meniscus height (TMH) and IOP with NCT and GAT were measured. In each patient, a comparison of IOP readings between the eye with lower TMH and the contralateral eye with higher TMH was evaluated. The TMH was graded as follows: grade 1 (low): TMH < 0.2 mm; grade 2 (moderate): 0.2 mm ≤ TMH < 0.6 mm; grade 3 (high): TMH ≥ 0.6 mm. Subsequently, a comparison of IOP readings among eyes with low, moderate, and high TMH was also performed. A total of 120 eyes of 60 patients were enrolled. When comparing the two eyes of a patient, the eye with higher TMH showed higher NCT readings and larger difference in IOP readings between the two tonometries than the eye with lower TMH (P < 0.001 and P < 0.001, respectively). When TMH was classified into grades according to the degree, the high TMH eyes showed higher NCT readings than did the low and moderate TMH eyes (P < 0.001 and P = 0.001, respectively). In addition, the high TMH eyes showed a larger difference in IOP readings between the two tonometries than did the low and moderate TMH eyes (P < 0.001 and P < 0.001, respectively). Eyes with higher TMH showed higher NCT readings and a larger difference in IOP between the two tonometries (NCT and GAT) than those with lower TMH. In patients with tearing, the NCT value may be inaccurate, so it is necessary to measure the GAT.Keywords
Funding Information
- VHS Medical Center Research Grant (VHSMC 18033)
This publication has 17 references indexed in Scilit:
- Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometerClinical Ophthalmology, 2012
- IOP Measured by Dynamic Contour Tonometry Correlates With IOP Measured by Goldmann Applanation Tonometry and Non-contact Tonometry in Japanese IndividualsJournal of Glaucoma, 2012
- THE NONCONTACT TONOMETERActa Ophthalmologica, 2009
- Quantitative Evaluation of Tear Meniscus Height From Fluorescein PhotographsCornea, 2007
- The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometryBritish Journal of Ophthalmology, 2005
- Tear meniscus level as an indicator of nasolacrimal obstructionOphthalmology, 2005
- Sources of error with use of Goldmann-type tonometersSurvey of Ophthalmology, 1993
- BIOMETRIC CORRELATIONS OF CORNEAL THICKNESSActa Ophthalmologica, 1975
- APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESSActa Ophthalmologica, 1975
- A NEW TONOMETER SYSTEMOptometry and Vision Science, 1972