Effects of high relative humidity and dry purging on VOCs obtained during breath sampling on common sorbent tubes
Authors:
Open Access
- 16 July 2020
- journal article
- research article
- Published by IOP Publishing in Journal of Breath Research
- Vol. 14 (4) , 046006
Abstract
Offline breath analysis by thermal desorption gas chromatography mass spectrometry (TD-GC-MS) requires the use of sorbent traps to concentrate and store volatile compounds. The selection of which sorbent to use and best practices for managing high relative humidity are important considerations to allow for reproducible, untargeted, biomarker discovery in water saturated breath samples. This work aims to assess three commonly used sorbent materials for their use in breath volatile sampling and determine how the high relative humidity inherent in such samples effects the capture of volatile compounds of interest. TenaxGR, TenaxTA/Carbograph1TD and TenaxTA/Carbograph5TD tubes were selected as they are the most commonly used sorbents in the breath sampling literature. The recovery of 29 compounds in a standard mix loaded using high humidity gas was tested for each sorbent and compared to loading in dry gas. Water retention and dry purge rates were determined for each sorbent for 500 ml and 1000 ml breath collections. Finally, breath samples were collected simultaneously on to each sorbent type using the ReCIVA and analysed by TD-GC-MS. All three sorbents exhibited acceptable reproducibility when loaded with the standard mix in dry gas (RSD < 10%). Loading the standard mix in humid gas led to reduced recovery of compounds based on their chemical properties. Dry purging performance for each sorbent material was assessed and was shown to be 1.14, 1.13 and 0.89 mg H2O min−1 for TenaxGR, TenaxTA/Carbograph1TD and TenaxTA/Carbograph5TD respectively when flushed with 50 ml min−1 of N2. A comparison of breath profiles on different sorbents showed differences in background artefacts (sulfur dioxide, cyclopenten-1-one and 3-nonene) and endogenous breath compounds (2-methyl-furan and furfural). This work demonstrates that high relative humidity during sampling reduces the ability of sorbent tubes to capture volatile compounds and could impact method detection limits during breath sampling. Sufficient water to impair accurate analysis was retained on all tubes. Minimal differences were observed between sorbent materials when used to sample breath, however, suggestions are provided for sorbent selection for future studies.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/M507490/1)
This publication has 30 references indexed in Scilit:
- A nanomaterial-based breath test for distinguishing gastric cancer from benign gastric conditionsBritish Journal of Cancer, 2013
- Predicting blood:air partition coefficients using basic physicochemical propertiesRegulatory Toxicology and Pharmacology, 2012
- Sorbent-based sampling methods for volatile and semi-volatile organic compounds in air. Part 2. Sorbent selection and other aspects of optimizing air monitoring methodsJournal of Chromatography A, 2010
- Spectral relative standard deviation: a practical benchmark in metabolomicsThe Analyst, 2008
- Impact of reaction products from building materials and furnishings on indoor air quality—A review of recent advances in indoor chemistryAtmospheric Environment, 2007
- Dry purge for the removal of water from the solid sorbents used to sample volatile organic compounds from the atmospheric airThe Analyst, 2000
- Anthropogenic volatile organic compounds in ambient air and natural waters: a review on recent developments of analytical methodology, performance and interpretation of field measurementsJournal of Chromatography A, 1999
- Monitoring VOCs in Air Using Sorbent Tubes Followed by Thermal Desorption-Capillary GC Analysis: Summary of Data and Practical GuidelinesJournal of the Air & Waste Management Association, 1997
- Comparison of Tenax Ta and Carbotrap for sampling and analysis of volatile organic compounds in airAtmospheric Environment. Part B. Urban Atmosphere, 1991
- Comparative assessment of the artefact background on thermal desorption of tenax GC and tenax TAJournal of Chromatography A, 1986