Clearance, Distribution Volume, and Dialyzer Mass Area Transport Coefficient of Glucose in Whole Blood
Authors:
- Daniel SchneditzDSDaniel Schneditz
- Edda ZierlerEZEdda Zierler
- Elisabeth MartinelliEMElisabeth Martinelli
- Roza Czabak-GarbaczRCRoza Czabak-Garbacz
- Mark HoehleinMHMark Hoehlein
- 1 March 2012
- journal article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 58 (2) , 137-142
Abstract
The extracorporeal transport of glucose was studied to determine the dialyzer mass transfer coefficient K(0)A for glucose in whole blood under conditions of glucose delivery and glucose removal. Glucose was removed from blood or delivered to blood using glucose-free dialysate or dialysate with a glucose concentration of 200 mg/dl (11.1 mmol/L). FX8 dialyzers (Fresenius Medical Care, Bad Homburg, Germany) were studied at constant dialysate flow Q(d) (500 ml/min) and variable blood flows Q(b) (200, 300, and 400 ml/min) under countercurrent flow conditions in a series of laboratory bench studies. Glucose clearance K(d) and glucose distribution volume flow rate Q(e) were determined from glucose mass balance. In 32 studies done with bovine blood at different hematocrit levels glucose was calculated to distribute in plasma water and to be excluded from red cell water when passing the FX8 dialyzer. The dialyzer mass transfer area coefficient K(0)A for glucose computed from Q(e), Q(d), and K(d) was 301.6 ± 45.2 ml/min and not different between modes of glucose delivery or glucose removal but lower than expected from the diffusivity of glucose estimated for aqueous solutions.Keywords
This publication has 16 references indexed in Scilit:
- Effect of changes in the intravascular volume during hemodialysis on blood viscoelasticityIndian Journal of Nephrology, 2011
- Fatigue in Hemodialysis Patients With and Without Diabetes: Results From a Randomized Controlled Trial of Two Glucose-Containing DialysatesDiabetes Care, 2010
- Intracorporeal Glucose Disposal During Hemodialysis After a Standardized Glucose LoadASAIO Journal, 2010
- A Quality Improvement Model for Optimizing Care of the Diabetic End-Stage Renal Disease PatientSeminars in Dialysis, 2010
- Glycemic Control and Burnt‐Out Diabetes in ESRDSeminars in Dialysis, 2010
- Is There a Role for Intradialytic Parenteral Nutrition? A Review of the EvidenceAmerican Journal of Kidney Diseases, 2010
- Rate of creatinine equilibration in whole bloodHemodialysis International, 2009
- The mechanism of hypoglycemia caused by hemodialysisClinical Nephrology, 2004
- RENAL RESEARCH INSTITUTE SYMPOSIUM: Effective Diffusion Volume Flow Rates (Qe) for Urea, Creatinine, and Inorganic Phosphorous (Qeu, Qecr, QeiP) During HemodialysisSeminars in Dialysis, 2003
- Low‐potassium and glucose‐free dialysis maintains urea but enhances potassium removalNephrology Dialysis Transplantation, 2001