A 100-MHz ultrasonic transducer array using ZnO thin films
Authors:
- K. SugawaraKSK. Sugawara
- H. TakeuchiHTH. Takeuchi
- 1 March 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 42 (2) , 316-324
Abstract
A 100-MHz ultrasonic linear transducer array made from a piezoelectric zinc oxide thin film on a sapphire substrate was developed and evaluated. Epitaxial, high-acoustic quality 10-/spl mu/m-thick ZnO film layers were produced by rf-magnetron sputter deposition onto a (111)-oriented gold film (with a chromium adhesion layer) that was vacuum-evaporated onto a (0001) sapphire surface. We found that, in well-oriented growth of gold, it is important to control the chromium sublayer thickness (less than 5 nm). An array was constructed by photolithography with an appropriate etch. V-shaped grooves between adjacent elements were formed by using an anisotropic etchant (HCl and HNO/sub 3/-based) that preferentially etched the c-plane of ZnO. Typical array elements were 90 /spl mu/m wide, 3.2 mm long, and 10 /spl mu/m thick, and the pitch of an array was typically 100 /spl mu/m. Our fine uniform array resulted in uniform ultrasonic response of individual elements throughout the array. For a 32-element array, the ultrasound beam in the azimuth plane in water could be electronically focused in the 100 MHz range to obtain a half-amplitude width of 60 /spl mu/m at the focal depth, agreeing well with theoretical predictions. Besides the use demonstrated with this present transducer, piezoelectric thin films should also lead to fabrication of various other kinds of ultrasonic transducers that can operate at high frequencies and should provide opportunities for miniaturizing transducers and making integrated ultrasonic devices.<>Keywords
This publication has 20 references indexed in Scilit:
- Scanning acoustic microscopyUltrasonics, 1993
- TfP210. Thin-film ZnO ultrasonic transducer arrays for operation at 100 MHzFerroelectrics, 1992
- What dermatologists should know about digital imagingJournal of the American Academy of Dermatology, 1991
- 100 MHz Band Ultrasonic Array Transducers Utilizing Epitaxially Grown ZnO films1991
- Subsurface Ultrasound Microscopic Imaging of the Intact EyeOphthalmology, 1990
- Planar-Structure Focusing Lens for Operation at 200 MHz and Its Application to the Reflection-Mode Acoustic Microscope1986
- Zinc Oxide Films for Acoustoelectric Device ApplicationsIEEE Transactions on Sonics and Ultrasonics, 1985
- Material Characterization by Line-Focus-Beam Acoustic MicroscopeIEEE Transactions on Sonics and Ultrasonics, 1985
- Penetration Depth of the Scanning Acoustic MicroscopeIEEE Transactions on Sonics and Ultrasonics, 1985
- Acoustic microscope—scanning versionApplied Physics Letters, 1974