EquiOx: A Prospective study of pulse oximeter bias and skin pigmentation in critically-ill adults
Preprint
Abstract
Rationale: Pulse oximeter performance may vary by skin pigmentation, but most data are retrospective with key limitations. Objective: To quantify pulse oximeter bias [mean difference between pulse oximeter oxygen saturation (SpO 2 ) and arterial blood functional oxygen saturation (SaO 2 )], and average root mean square error (A RMS ) and estimate adjusted effects of skin pigment on bias or A RMS in critically-ill adults Methods: Prospective single-center study of 631 ICU patients (2022–2024) directly observed SpO and SaO pairs. Skin pigment was assessed using the subjective Monk Skin Tone Scale and objective spectrophotometry measurement (Individual Typology Angle [ITA]). Adjusted effects were estimated with targeted maximum likelihood estimation. Main Results: Among 1,760 paired measurements from 631 critically-ill adults, median SaO 2 was 98% (IQR 96%, 99%) with 40 episodes of stable hypoxemia (SpO 2 2 systematically underestimated SaO 2 [median bias= −1.70 IQR (−2.84, −0.50)]. Bias was less negative in patients with darker skin (ITA <-30°) [−1.05 (−2.44, −0.10)] vs lighter skin (ITA >30°) [−2.01 (−3.34, −1.00)] and remained significantly different in adjusted analyses. A RMS was 3.87 (95% CI 3.25, 4.53) overall and 4.49 (95% CI 2.63, 7.07) in patients with darker skin. Ear probes performed worse than finger probes [bias: −2.20 vs. −1.60; A RMS : 4.90 vs. 2.70]. Conclusions: In this large ICU cohort, hypoxemia was rare, pulse oximeters systematically underestimated SaO, performance varied by skin pigment and probe site. Bias was less negative in patients with darker skin. Pulse oximeter inaccuracies in ICU patients may be more substantial than clinicians recognize. Key Points: Question: Does pulse oximeter bias in critically-ill adults differ by skin pigmentation? Findings: In this prospective, single-center study of 631 critically-ill adults, oximeter mean bias was negative for all patients but relatively less negative for patients with darker pigment. Bias variation by pigment was nonlinear and larger in ear than finger probes. Meaning: Pulse oximeter bias varies with skin pigmentation and may not always be positive in patients with darker pigmentation. Pulse oximeter inaccuracy may be larger than clinicians appreciate. Additional studies with multiple oximeter brands and more stable hypoxemia are needed to further understand bias variation with skin pigment.Keywords
This publication has 43 references indexed in Scilit:
- Race and Ethnicity in Biomedical ResearchJAMA, 2025
- In Real-World Studies, Comparisons of Predictive Values Are Confounded by PrevalenceChest, 2022
- Assessment of Racial and Ethnic Differences in Oxygen Supplementation Among Patients in the Intensive Care UnitJAMA Internal Medicine, 2022
- Racial and Ethnic Discrepancy in Pulse Oximetry and Delayed Identification of Treatment Eligibility Among Patients With COVID-19JAMA Internal Medicine, 2022
- Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019-2020Chest, 2022
- Analysis of Discrepancies Between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association With Organ Dysfunction and MortalityJAMA Network Open, 2021
- Reconsidering the Consequences of Using Race to Estimate Kidney FunctionJAMA, 2019
- Targeted Learning2011
- Research electronic data capture (REDCap)—A metadata-driven methodology and workflow process for providing translational research informatics supportJournal of Biomedical Informatics, 2009
- Accuracy of pulse oximetry in the intensive care unitIntensive Care Medicine, 2001