Tailoring cardiac resynchronization therapy using interventricular asynchrony. Validation of a simple model
Authors:
- Angelo AuricchioAAAngelo Auricchio
- Yinghong YuYYYinghong Yu
- Jiang DingJDJiang Ding
- Thierry PochetTPThierry Pochet
- Kevin VernooyKVKevin Vernooy
- Andrew KramerAKAndrew Kramer
- Julio SpinelliJSJulio Spinelli
- Frits W. PrinzenFPFrits W. Prinzen
- 1 March 2006
- journal article
- clinical trial
- Published by American Physiological Society in American Journal of Physiology-Heart and Circulatory Physiology
- Vol. 290 (3) , H968-H977
Abstract
This study explores the use of interventricular asynchrony (interVA) for optimizing cardiac resynchronization therapy (CRT), an idea emerging from a simple pathway model of conduction in the ventricles. Measurements were performed in six dogs with chronic left bundle branch block (LBBB) and in 29 patients of the Pacing Therapies for Congestive Heart Failure (PATH-CHF)-I study. In the dogs, intraventricular asynchrony (intraVA) was determined using left ventricular (LV) endocardial activation maps. In dogs and patients, the maximum rate of rise of LV pressure (LV dP/d tmax) and the pulse pressure (PP) and interVA [time delay between upslope of LV and right ventricular (RV) pressure curves] were measured during LV, RV, and biventricular (BiV) pacing with various atrioventricular (AV) delays. Measurements in the canine hearts supported the pathway model in that optimal resynchronization occurred at ∼50% reduction of intraVA and at an interVA value halfway that during LBBB and LV pacing. In patients with significant hemodynamic response during pacing ( n = 22), intrinsic interVA and interVA at peak improvement (interVAp) varied widely between patients (from −83 to −15 ms and from −42 to +31 ms, respectively). However, the model predicted individual interVAp accurately (SD of ±6 ms and ±12 ms for LV dP/d tmax and PP, respectively). At equal interVA, LV and BiV pacing produced equal hemodynamic response, but in 11 of 22 responders, BiV pacing reduced interVA insufficiently to reach the maximum hemodynamic response. LV pacing at short AV delay proved to result in better hemodynamics than predicted by the model, indicating that additional factors determine hemodynamics during LV preexcitation. Guided by a simple pathway model, interVA measurements accurately predict optimal hemodynamic performance in individual CRT patients.Keywords
This publication has 32 references indexed in Scilit:
- Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacingJACC, 2004
- Cardiac-Resynchronization Therapy with or without an Implantable Defibrillator in Advanced Chronic Heart FailureNew England Journal of Medicine, 2004
- Characterization of Left Ventricular Activation in Patients With Heart Failure and Left Bundle-Branch BlockCirculation, 2004
- Cardiac resynchronization therapy can reverse abnormal myocardial strain distribution in patients with heart failure and left bundle branch blockJACC, 2003
- Cardiac Resynchronization in Chronic Heart FailureNew England Journal of Medicine, 2002
- Effect of Pacing Chamber and Atrioventricular Delay on Acute Systolic Function of Paced Patients With Congestive Heart FailureCirculation, 1999
- Evaluation of Different Ventricular Pacing Sites in Patients With Severe Heart FailureCirculation, 1997
- Hemodynamic importance of preserving the normal sequence of ventricular activation in permanent cardiac pacingAmerican Heart Journal, 1995
- Functional abnormalities in isolated left bundle branch block. The effect of interventricular asynchrony.Circulation, 1989
- Physiology of Canine Intraventricular Conduction and Endocardial ExcitationCirculation Research, 1972