Experimental His‐Bundle Pacing: Histopathological and Electrophysiological Examination

Authors:

Abstract
His‐bundle pacing gives a more physiological ventricular contraction in comparison to right ventricular apical pacing. However the problems of lead fixation and stability of long‐term His‐bundle pacing are yet unsolved. We used six adult beagles, in which a screw‐in lead was anchored in the His‐bundle region for observation of the pacing conditions and histopathologic changes of the conduction system over the course of 2 months. In the results, a satisfactory fixation was obtained using a conventional screw‐in lead and no histological influence on the conduction system was observed. The pacing threshold at the time of implantation was 1.15 ± 0.69V (3.23 ± 3.08 mA) in the pulse width of 0.5 ms. R wave amplitude, the impedance and slew rate were 7.28 ± 2.04 mV, 409 ± 102 Ohm, and 0.65 ± 0.41 V/s, respectively. Two months later, these parameters changed to 2.83 ± 1.06 V (10.4 ± 5.71 mA), 5.63 ± 1.62 mV, 310 ± 71.3 Ohm, and 0.49 ± 0.22 V/s, respectively. These results suggest the feasibility of clinical application of permanent His‐bundle pacing.