Comparative study of minimal fresh gas flow used in Lack-Plus and Lack’s circuit in spontaneously breathing anesthetized adults
Open Access
- 1 November 2016
- journal article
- research article
- Published by Taylor & Francis in Medical Devices: Evidence and Research
- Vol. ume 9, 395-399
Abstract
Background: The Lack’s circuit is a co-axial Mapleson A breathing system commonly used in spontaneously breathing anesthetized adults but still requires high fresh gas flow (FGF). The Lack-Plus circuit was invented with the advantage of lower FGF requirement. The authors compared the Lack-Plus and Lack’s circuit for the minimal FGF requirement with no rebreathing in spontaneously breathing anesthetized adults. Methods: This was a randomized crossover study. We enrolled 24 adult patients undergoing supine elective surgery, with a body mass index ≤30 kg/m2 and an American Society of Anesthesiologists physical status I–II. They were randomly allocated to group 1 (LP-L) starting with Lack-Plus then switching to Lack’s circuit or group 2 (L-LP) (with the reverse pattern). After induction and intubation, anesthesia was maintained with 50% N2O/O2 and desflurane (4%–6%) plus fentanyl titration to maintain an optimal respiratory rate between 10 and 16/min. Starting with the first circuit, all the patients were spontaneously breathing with a FGF of 4 L/min for 10 min, gradually decreased by 0.5 L/min every 5 min until FGF was 2.5 L/min. End-tidal CO2, inspired minimum CO2 (ImCO2), mean arterial pressure, and oxygen saturation were recorded until rebreathing (ImCO2 >0 mmHg) occurred. The alternate anesthesia breathing circuit was used and the measurements were repeated. Results: The respective minimal FGF at the point of rebreathing for the Lack-Plus and Lack’s circuit was 2.7±0.8 and 3.3±0.5 L/min, respectively, p2 was 1.5±2.0 and 4.2±2.6 mmHg, respectively, pConclusion: The Lack-Plus circuit can be used safely and effectively, and it requires less FGF than Lack’s circuit in spontaneously breathing anesthetized adults.Keywords
This publication has 7 references indexed in Scilit:
- Comparison of minimal fresh gas requirements of baby enclosed afferent reservoir and Jackson Rees anesthetic circuit for general anesthesia in spontaneously breathing children.2014
- A clinical determination of optimal fresh gas flow in a baby EAR circuit.2009
- An Enclosed Efferent Afferent Reservoir System: The MaximaAnaesthesia and Intensive Care, 1995
- ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEMSBritish Journal of Anaesthesia, 1988
- The Lack, Magill and Bain Anaesthetic Breathing Systems: A Direct Comparison in Spontaneously-Breathing Anaesthetized AdultsJournal of the Royal Society of Medicine, 1982
- Fresh Gas Original Investigations of the Magill CircuitAnesthesiology, 1968
- Fresh Gas Flow and Rebreathing in the Magill Circuit with Spontaneous RespirationProceedings of the Royal Society of Medicine, 1967