Abstract
Objective. To test whether or not the modified design of the test implant (intended to increase primary stability) has an equivalent effect on MBL compared to the control. Methods. Forty patients were randomly assigned to receive test or control implants to be installed in identically dimensioned bony beds. Implants were radiographically monitored at installation, at prosthetic delivery, and after one year. Treatments were considered equivalent if the 90% confidence interval (CI) for the mean difference (MD) in MBL was in between −0.25 and 0.25 mm. Additionally, several soft tissue parameters and patient-reported outcome measures (PROMs) were evaluated. Linear mixed models were fitted for each patient to assess time effects on response variables. Results. Thirty-three patients (21 males, 12 females; years old) with 81 implants (47 test, 34 control) were available for analysis after a mean observation period of months (3 dropouts, 3 missed appointments, and 1 missing file). The adjusted MD in MBL after one year was −0.13 mm (90% CI: −0.46–0.19; test group: −0.49; control group: −0.36; ). Conclusion. Both implant systems can be considered successful after one year of observation. Concerning MBL in the presented setup, equivalence of the treatments cannot be concluded. Registration. This trial is registered with the German Clinical Trials Register (ID: DRKS00007877).
Keywords
This publication has 26 references indexed in Scilit:
- Reporting of Noninferiority and Equivalence Randomized TrialsJAMA, 2012
- Magnitude and Direction of Mechanical Stress at the Osseointegrated Interface of the Microthread ImplantJournal of Periodontology, 2011
- Effect of conical configuration of fixture on the maintenance of marginal bone level: preliminary results at 1 year of functionClinical Oral Implants Research, 2010
- Comparative Analysis of Peri‐Implant Marginal Bone Loss Based on Microthread Location: A 1‐Year Prospective Study After LoadingJournal of Periodontology, 2009
- Effect of Microthreads and Platform Switching on Crestal Bone Stress Levels: A Finite Element AnalysisJournal of Periodontology, 2008
- Effect of microthread on the maintenance of marginal bone level: a 3‐year prospective studyClinical Oral Implants Research, 2007
- Living with cracks: Damage and repair in human boneNature Materials, 2007
- Tissue Characteristics at Microthreaded Implants: An Experimental Study in DogsClinical Implant Dentistry and Related Research, 2006
- Concise Review Clinical: Equivalence Testing with Dental Clinical TrialsJournal of Dental Research, 2001
- Outcomes and Their Measurement in Clinical Trials of Endosseous Oral ImplantsAnnals of Periodontology, 1997