Accuracy of a New Low-flow Sidestream Capnography Technology in Newborns: A Pilot Study
Authors:
Open Access
- 1 April 2002
- journal article
- research article
- Published by Springer Nature in Journal of Perinatology
- Vol. 22 (3) , 219-225
Abstract
OBJECTIVE: To evaluate the accuracy of a new low-flow sidestream capnography technology and analyze components of the capnogram in mechanically ventilated newborns with and without pulmonary disease. METHODS: Twenty patients were prospectively identified. Eligible infants were mechanically ventilated and had an indwelling arterial catheter. Two groups were identified: newborns who were receiving mechanical ventilation for pulmonary diseases, and newborns who were receiving postoperative mechanical ventilation for nonpulmonary conditions. End-tidal CO2 (PetCO2) was measured for 1-minute pre- and post-arterial blood sampling, and PetCO2 and PaCO2 were compared for each patient. Eight quantitative waveform parameters were also measured on all patients. RESULTS: Newborns in the pulmonary group (n=13) (persistent pulmonary hypertension of the newborn/meconium aspiration syndrome, respiratory distress syndrome, pneumonia) and newborns in the control group (n=7) were matched for birth weight, gestational age, and postnatal age. PetCO2–PaCO2 Gradient values were higher in the pulmonary group (7.4±3.3 mm Hg) than controls (3.4±2.4 mm Hg). Four waveform parameters (ascending slope, alveolar angle, alpha angle, descending angle) were identified, which independently differentiated patients with pulmonary disease from controls. CONCLUSIONS: Low-flow capnography with Microstream technology accurately measured alveolar CO2 in newborns without pulmonary disease, as demonstrated by normal PetCO2–PaCO2 gradients. The measured PetCO2–PaCO2 gradient, as expected, was significantly higher in newborns with pulmonary disease. We also identified four quantitative waveform parameters that may be useful in differentiating between mechanically ventilated newborn patients with and without lung disease.Keywords
This publication has 23 references indexed in Scilit:
- Randomized, Controlled Trial of Low-dose Inhaled Nitric Oxide in the Treatment of Term and Near-term Infants With Respiratory Failure and Pulmonary HypertensionPediatrics, 1999
- Trend in mortality from respiratory distress syndrome in the United States, 1970-1995The Journal of Pediatrics, 1999
- A prospective randomized comparison of conventional mechanical ventilation and very early high frequency oscillatory ventilation in extremely premature newborns with respiratory distress syndromeIntensive Care Medicine, 1999
- Improved Oxygenation in a Randomized Trial of Inhaled Nitric Oxide for Persistent Pulmonary Hypertension of the NewbornPediatrics, 1997
- Utility of the Expiratory Capnogram in the Assessment of BronchospasmAnnals of Emergency Medicine, 1996
- Expiratory capnography in asthma: evaluation of various shape indicesEuropean Respiratory Journal, 1994
- Technical and clinical aspects of capnography in neonatesJournal of Medical Engineering & Technology, 1991
- End-tidal Pco2 Monitoring in Infants and Children Ventilated with Either a Partial Rebreathing or a Non-rebreathing CircuitAnesthesiology, 1987
- Accuracy of expiratory carbon dioxide measurements using the coaxial and circle breathing circuits in small subjectsJournal of Clinical Monitoring and Computing, 1985
- Can we trust end-tidal carbon dioxide measurements in infants?Journal of Clinical Monitoring and Computing, 1985