Validation of VASO cerebral blood volume measurement with positron emission tomography
Open Access
- 2 December 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 65 (3) , 744-749
Abstract
Cerebral blood volume (CBV) has been shown to be an important biomarker in a number of neurological disorders and in the quantitative interpretation of functional MRI. One approach to determine CBV in humans is vascular‐space‐occupancy MRI, and this technique has been applied to the studies of brain glioma, Schizophrenia, and Alzheimer's disease. However, validation of this technique with a gold standard method has not been reported. In this study, we compared vascular‐space‐occupancy MRI with a radiotracer‐based positron emission tomography technique in a group of healthy subjects. It was found that regional CBV measured with vascular‐space‐occupancy MRI was highly correlated with that of the positron emission tomography data (R = 0.79 ± 0.10, N = 8). Furthermore, absolute CBV values quantified by vascular‐space‐occupancy were also in excellent agreement with those by positron emission tomography (slope = 1.00 ± 0.15). Because of the differences in the labeling principles between the two modalities, systematic CBV differences were observed in large vessel and ventricle regions. Magn Reson Med, 2011.Keywords
Funding Information
- NIH (R21 NS054916, R01 MH084021, R21 EB007821)
- American Heart Association (0865003F)
This publication has 20 references indexed in Scilit:
- Cerebral blood volume in Alzheimer's disease and correlation with tissue structural integrityNeurobiology of Aging, 2010
- Perfusion Deficit in Schizophrenia and Correlation with Psychopathological SymptomsNeuroImage, 2009
- On the measurement of absolute cerebral blood volume (CBV) using vascular‐space‐occupancy (VASO) MRIMagnetic Resonance in Medicine, 2008
- Predicting Grade of Cerebral Glioma Using Vascular-Space Occupancy MR ImagingAmerican Journal of Neuroradiology, 2008
- Evaluation of human brain tumor heterogeneity using multiple T1‐based MRI signal weighting approachesMagnetic Resonance in Medicine, 2008
- Cerebral Vascular Mean Transit Time in Healthy Humans: A Comparative Study with PET and Dynamic Susceptibility Contrast-Enhanced MRIJournal of Cerebral Blood Flow & Metabolism, 2007
- Absolute CBF and CBV measurements by MRI bolus tracking before and after acetazolamide challenge: Repeatabilily and comparison with PET in humansNeuroImage, 2005
- Hemodynamic segmentation of MR brain perfusion images using independent component analysis, thresholding, and Bayesian estimationMagnetic Resonance in Medicine, 2003
- Absolute Quantification of Cerebral Blood Flow with Magnetic Resonance, Reproducibility of the Method, and Comparison with H215O Positron Emission TomographyJournal of Cerebral Blood Flow & Metabolism, 2002