The use of gel dosimetry to measure the 3D dose distribution of a90Sr/90Y intravascular brachytherapy seed
Authors:
- G Massillon-JlGMG Massillon-Jl
- R MinnitiRMR Minniti
- M G MitchMMM G Mitch
- M J MaryanskiMMM J Maryanski
- C G SoaresCSC G Soares
- 25 February 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (6) , 1661-1672
Abstract
Absorbed dose distributions in 3D imparted by a single (90)Sr/(90)Y beta particle seed source of the type used for intravascular brachytherapy were investigated. A polymer gel dosimetry medium was used as a dosemeter and phantom, while a special high-resolution laser CT scanner with a spatial resolution of 100 microm in all dimensions was used to quantify the data. We have measured the radial dose function, g(L)(r), observing that g(L)(r) increases to a maximum value and then decreases as the distance from the seed increases. This is in good agreement with previous data obtained with radiochromic film and thermoluminescent dosemeters (TLDs), even if the TLDs underestimate the dose at distances very close to the seed. Contrary to the measurements, g(L)(r) calculated through Monte Carlo simulations and reported previously steadily decreases without a local maximum as a function of the distance from the seed. At distances less than 1.5 mm, differences of more than 20% are observed between the measurements and the Monte Carlo calculations. This difference could be due to a possible underestimation of the energy absorbed into the seed core and encapsulation in the Monte Carlo simulation, as a consequence of the unknown precise chemical composition of the core and its respective density for this seed. The results suggest that g(L)(r) can be measured very close to the seed with a relative uncertainty of about 1% to 2%. The dose distribution is isotropic only at distances greater than or equal to 2 mm from the seed and is almost symmetric, independent of the depth. This study indicates that polymer gel coupled with the special small format laser CT scanner are valid and accurate methods for measuring the dose distribution at distances close to an intravascular brachytherapy seed.Keywords
This publication has 23 references indexed in Scilit:
- Verification and uniformity control of doses for 90 Sr/ 90 Y intravascular brachytherapy sources using radiochromic film dosimetryJournal of Medical Physics, 2008
- Dose calculation formalisms and consensus dosimetry parameters for intravascular brachytherapy dosimetry: Recommendations of the AAPM Therapy Physics Committee Task Group No. 149Medical Physics, 2007
- Experimental determination of the radial dose function of Sr90∕Y90 IVBT sourcesMedical Physics, 2006
- Characterization of a high-dose-rate90Sr–90Y source for intravascular brachytherapy by using the Monte Carlo code PENELOPEPhysics in Medicine & Biology, 2002
- Long-Term Efficacy of Intracoronary Irradiation in Inhibiting In-Stent RestenosisCirculation, 2001
- Beta versus gamma for catheter-based intravascular brachytherapy: dosimetric perspectives in the presence of metallic stents and calcified plaquesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2000
- Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapyMedical Physics, 1999
- Endovascular β-Radiation to Reduce Restenosis After Coronary Balloon AngioplastyCirculation, 1998
- Nuclear Data Sheets for A = 90Nuclear Data Sheets, 1997
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993