Leg-length discrepancy measured by ultrasonography
Authors:
Open Access
- 1 January 1991
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 62 (2) , 121-124
Abstract
Leg-length discrepancy was determined by real-time ultrasonography (ultrasound) in 45 patients, and the measurements were compared with those of erect-posture radiography. A special device for holding and moving the ultrasound transducer was constructed, and the leg length was measured as the highest level of the femoral head in the standing position. The correlation coefficient r between ultrasound and radiography was 0.94, the mean difference was -1.9 mm, and the limits of agreement (mean ° 2 SD) were -9.1 to 5.3 mm. The mean difference between examiners 1 and 2 was 1.7 mm, and the 95 percent confidence interval was ± 7 mm. We conclude that leg-length discrepancy can be reliably determined by ultrasound, although the accuracy is less than that obtained by radiographic methods. Because ultrasound is not limited by radiation hazards, our technique can be used for clinical screening.Keywords
This publication has 5 references indexed in Scilit:
- Lower limb-length discrepancy.Journal of Bone and Joint Surgery, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Low-Back Pain Associated With Leg Length InequalitySpine, 1981
- UNEQUAL LEG LENGTH: AN ACCURATE METHOD OF DETECTION AND SOME CLINICAL RESULTSRheumatology, 1972
- Short-leg SyndromeBMJ, 1960