2008 Otto Aufranc Award: Component Design and Technique Affect Cement Penetration in Hip Resurfacing

Authors:

Abstract
Ue on cement penetration. Six retrieved femoral heads were resurfaced for each implant (BHR®, ASR®, Conserve Plus®, DuROM®, ReCAP®) using the manufacturers’ recommendations for implantation. In addition, the BHR was implanted using the Conserve Plus® high-viscosity cementation technique, “BHR/hvt,” and vice versa for the Conserve, “Conserve/lvt.” The average cement penetration was highest with BHR (65.62% ± 15.16%) compared with ASR® (12.25% ± 5.12%), Conserve Plus® (19.43% ± 5.28%), DuROM® (17.73% ± 3.96%), and ReCAP® (26.09% ± 5.20%). Cement penetration in BHR/hvt remained higher than all other implants equaling 36.7% ± 6.6%. Greater femoral component design clearance correlated with cement mantle thickness. Femoral component design in hip resurfacing plays a major role in cement penetration....

This publication has 46 references indexed in Scilit: