An analog front-end circuit with spike detection for implantable neural recording system design
- 1 August 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 15483746,p. 881-884
Abstract
A low power, low noise implantable neural recording interface for use in a power harvested inductive coupling system is presented in this paper. A two stage neural amplifier, selective reference thresholder and a serial communication interface form the neural data logger. The midband gain of neural amplifier is 59.1dB with a 3dB bandwidth from 0.45 to 8 kHz. Input-referred noise and total power consumption are 8.7μVrms and 3.19μW respectively at less than 1 V. The thresholder detects/quantizes the amplified neural signal at sampling rate of 20 kHz with four code selectable reference voltages. The 8 bit serial digits are transmitted via light emitting diode (LED) to a photo detector interface. The chip is designed and fabricated in 0.18-μm process. The simulation result demonstrates the neural recording interface's suitability for instu neutral activity recording.Keywords
This publication has 6 references indexed in Scilit:
- A 2μW digital baseband core for wireless Micro-Neural-Interface in 0.18μm CMOS2013
- A 700mV low power pipeline ADC using a novel common mode feedback circuit and offset cancellation technique2012
- A Battery-Powered Activity-Dependent Intracortical Microstimulation IC for Brain-Machine-Brain InterfaceIEEE Journal of Solid-State Circuits, 2011
- A 20µW neural recording tag with supply-current-modulated AFE in 0.13µm CMOS2010
- A Low-Power Integrated Circuit for a Wireless 100-Electrode Neural Recording SystemIEEE Journal of Solid-State Circuits, 2006
- A 100-channel system for real time detection and storage of extracellular spike waveformsJournal of Neuroscience Methods, 1999